7.23
Second analog output option low value
Display:
REC_ Ch2
Range:
Any display value
Default Value:
0
See
REC_
function
for description of operation.
7.24
Second analog output option high value
Display:
REC~ Ch2
Range:
Any display value
Default Value:
1000
See
REC~
function
for description of operation.
7.25
Display rounding
Display:
drnd
Range:
1
to
5000
Default Value:
1
Displays and sets the display rounding value for the rate/frequency display only. This value may
be set to 1 - 5000 displayed units. Display rounding is useful for reducing the instrument resolution
without loss of accuracy in applications where it is undesirable to display to a fine tolerance. To
set the display rounding value go to the
drnd
function and use the
^
or
v
push buttons to set
the required value then press
F
to accept this selection.
Example:
If set to
10
the rate/frequency display values will change in multiples of 10 only i.e. display moves
from
10
to
20
to
30
etc.
7.26
Digital filter
Display:
FLtr
Range:
0
to
8
Default Value:
2
Displays and sets the digital filter value.
Digital filtering uses a weighted average method of
determining the display value and is used for reducing display value variation due to short term
interference. The digital filter range is selectable from
0
to
8
, where
0
= none and
8
= most
filtering. Use
^
or
v
at the
FLtr
function to alter the filter level if required. Note that the
higher the filter setting the longer the display may take to reach its final value when the input is
changed, similarly the relay operation and any output options will be slowed down when the filter
setting is increased. To set the digital filter value go to the
FLtr
function and use the
^
or
v
push buttons to set the required value then press
F
to accept this selection.
PM4TRMAN-2.7-0
37 of