then the display will reset when the display value reaches the pass value (set at
A1PS
) and the
relay will activate and will remain activated for the time set at the
A1Pt
function.
7.65
Default display
Display:
dFIt dISP
Range:
RAtE
,
PERd
or
totI
Default Value:
RAtE
Seen only in
both
mode. Sets the display default to either total or rate (or period if
dISP
is
set to
PERd
). The display will always revert to the default display on power up but can be forced
to the alternate display via the
P
button or remote input (see
P.but
and
R.INP
functions). If
the alternate display is being viewed an identifier will flash briefly every 8 seconds approx. e.g. if
the rate is the default display and the display is now toggled to show total the message
totI
will
flash briefly every 8 seconds.
7.66
Set display operation
Display:
SEt OPER
Range:
S.Prd
,
PERd
,
FREQ
,
totL
or
both
Default Value:
FREQ
Displays and sets the selected operating mode, e.g. select
totL
for totaliser operation. Options
are:
S.Prd
- Not applicable to this instruction manual
PERd
- Not applicable to this instruction manual
both
- Frequency and total - allows toggling between rate and total display
totL
- Total/counter display
FREQ
- Frequency/rate display
7.67
Baud rate for optional serial communications
Display:
bAUd RAtE
Range:
300, 600, 1200, 2400, 4800, 9600, 19.2
or
38.4
Default Value:
9600
Set baud rate - seen only with serial output option. Refer to the separate “PM4 Panel Meter
Optional Output Addendum” booklet supplied when optional outputs are fitted.
Select from
300, 600, 1200, 2400, 4800, 9600, 19.2
or
38.4
baud. The baud rate should be set to
match the device being communicated with.
PM4TRMAN-2.7-0
51 of