4.88
Data string timeout
Display:
t.out
Range:
0.0
to
10.0
Default Value:
1.0
This function allows the user to set a timeout value for the data stream. Valid times are
0.0
to
10.0
seconds, a setting of
0.0
disables the timeout. The timeout will cause the current data stream to be
ignored if the time gap between characters in the stream exceeds the
t.out
value. This function helps
to prevent false displays when the data stream is interrupted.
4.89
P
button function
Display:
P.but
Range:
NONE, Hi, Lo, HILo, tARE
or
ZERO
Default Value:
NONE
P
button function - The
P
button function can only be used when the
COdE
function is set to
UAL
. The
P
button may be set to operate some of the remote input functions. With the tare and zero functions, to
prevent accidental operation, the
P
button must be held pressed for 2-3 seconds before the display will
tare or zero, momentary operation of the tare function will cause the gross value to be displayed, preceded
by the message
GROS
. If both the remote input and
P
button function are operated simultaneously the
P
button will override the remote input. The functions below are as described in the
R.INP
function
below. Functions available are:
NONE, Hi, Lo, HiLo, tARE
or
ZERO
4.90
Remote input function
Display:
R.INP
Range:
NONE,P.HLd,d.HLd,Hi,Lo,HiLo,tARE,ZERO,SP.Ac,No.Ac
or
duLL
Default Value:
NONE
Remote input functions can only be used when the
CodE
function is set to
UAL
.
Remote input function - When these remote input terminals are short circuited, via a switch, relay,
keyswitch etc. the instrument will perform the selected remote input function. A message will flash to
indicate which function has been selected when the remote input pins are short circuited. The remote
input functions are as follows:
NONE
- no remote function required i.e. activating the remote input has no effect.
P.HLd
- peak hold. The display will show the peak value (highest positive value) only whilst the remote
input terminals are short circuited i.e. the display value can rise but not fall whilst the input
terminals are short circuited. The message
P.HLd
will appear briefly every 8 seconds whilst the
input terminals are short circuited to indicate that the peak hold function is active.
d.HLd
- display hold. The display value will be held whilst the remote input terminals are short circuited.
The message
d.HLd
will appear briefly every 8 seconds whilst the input terminals are short circuited
to indicate that the display hold function is active.
Hi
- peak memory. The peak value stored in memory will be displayed if the remote input terminals are
short circuited, if the short circuit is momentary then the display will return to normal measurement
after 20 seconds. If the short circuit is held for 2 to 3 seconds or the power is removed from the
instrument then the memory will be reset.
Lo
- valley memory. The minimum value stored in memory will be displayed. Otherwise operates in the
same manner as the
Hi
function described above.
LDRSMANADD-2.3-0
37 of