Manual mode:
1. Connect the power cord and turn on the power switch. When the switch is on, the
power LED lights up.
2. Set the auto/manual switch to the left side, the auto/manual LED lights off.
3. Next, slide the D2 switch or halogen switch to the right according to the purpose, and
one or both LED lights up.
4. When the shutter switch is set to the right and the LED lights up, the optical path opens
and it will be in measurement state. When the shutte switch is set to the left and the
LED lights off, the optical path is cut off and it will be setting as dark spectrum state.
Automatic mode:
1. Connect the spectrometer and the light source with the trigger cable.
2. Connect the spectrometer and the PC with USB cable.
3. Turn on the power switch of the light source and slide the auto/manual switch to the
right side, the LED lights up.
4
.
SpectraSmart controls D2, halogen and shutter operation. Check ON/OFF of each
switch by LED status.
Caution: Turning on the power switch before connecting the ligth source control trigger
cable, it does not work properly. Please be careful to turn on the light source only after
connecting the light source control trigger cable.
Figure 3-5: Connection of the light source control trigger cable