3.2. Considerations for the position of devices
In order to ensure sufficient space for an experiment, place the PC on the MED64-Entry amplifier (not essential). Pay attention so
as not to damage the housing of the MED64-Entry Amplifier with the legs of the PC. Place the display monitor next to the PC and
the MED Connector in front of the monitor. Do not place the power adaptor or other electronic devices close to the connection
terminal on the back of the amplifier.
Above is an image of the ideal positions of components. The user can perform an experiment smoothly when there
is a sufficient space next to or at the back of the laboratory desk with a power strip on the floor. If no perfusion
system is used, about 100 cm width is sufficient, otherwise a larger space is necessary.
3.3. Power source
Three power sources are necessary for the MED64-Entry Amplifier and the Data Acquisition PC System (the desktop PC and the
display monitor), and one additional power source is necessary to use the MED Thermo connector when it is used. Connect all the
cords to one power strip connected to a wall outlet with the ground terminal (do not use an outlet on the desktop rack because it
is often not grounded properly). Do not connect devices that are not necessary for the MED64-Entry to the power strip or to the
wall outlet to which the MED64-Entry and other components are connected. Place the power cable and the power adaptor at a
distance from the amplifier, the MED Connector and the 68-pin SCSI cable.
Display monitor
PC
MED64-Entry
Amplifier
Space for cables on the back of the amplifier
About 20 cm depth is necessary.
MED Connector
Mouse
Keyboard
PC
Display monitor
Power strip (on the floor)
Power adaptor (on the floor)
75 cm
120 cm
アンプ後面
Data acquisition device other
than PCIe-6343 (not connected).
Save after assignment
of the name.
Space for cables in front of the amplifier
About 15 cm depth is necessary.
Pump (when
perfusing)
Water bath
(when perfusing)
MED64-Entry Amplifier