15
In EP mode, two perpendicular cursors appear on the signal waveform chart. Use drag & drop to change the position on the X-axis
and calculate the minimum value of the raw data between the two perpendicular cursors to obtain the indicator of stability of
activity.
SPIKE mode (left) and ISI mode (right) of BASELINE STABILITY panel.
In SPIKE mode or ISI mode, one horizontal cursor (threshold) appears on the raw data chart. Use drag & drop to change the
position on the Y-axis, to count the frequency of the signal crossing with the threshold (SPIKE), or to calculate the interval between
crossings (ISI: inter-spike interval). The target of the measure is the cross point for rising when the threshold is on the positive side,
and the cross point for falling when the threshold is on the negative side.
The same as in the SIGNAL WAVEFORM panel, a menu to change the scope to display on an axis appears by right clicking on the
signal waveform chart. The value specified on the X-axis indicates the maximum value (s) on the X-axis. For the Y-axis, click the axis
label of the extreme value to change the scope of display by direct editing.
Other menus are the same as those available in the SIGNAL WAVEFORM panel.
The measure chart is similar to the signal waveform chart. Step Filter shows only the evoked response at either step as a time chart,
for data acquisition with Step 1 and 2 enabled. Copy Data stores the numerical data of the time chart on a clipboard. Moving the
mouse cursor over each plot of the chart shows the numerical information of that plot point.
Summary of Contents for MED64-Entry
Page 1: ...MED64 Entry Manual ...