OPERATION MANUAL
Repeat the treatment once or twice a day in compliance with dosages.
Keep me interval between sessions at least 8 hours.
Among persons with arterial hypertension, blood pressure should be monitored
in the
fi
rst 6 days of treatment just before session and 20 to 30 minutes a er treat-
ment. This is necessary to evaluate the magne c
fi
eld sensi vity and assess quality
of treatment.
In case the exposure results in the blood pressure
increase or decrease by 10
to 25 mm Hg
during the
fi
rst 6 days of treatment (Mode 3), reduce the treatment
dura on by 1/3 during the next session.
In case the exposure results in the blood pressure
increase or decrease by 10
to 25 mm Hg
during the
fi
rst 6 days of treatment (Mode 1), apply Mode 2 during
the next 3 days of treatment. From that me onwards, resume Mode 1 treatment.
In case the exposure results in the blood pressure
increase or decrease by more
than 25 mm Hg
consult with your local physiotherapist or a ending doctor prior
to the next session to adjust the treatment procedure. Keep a log by entering the
measurement data required to trace the me course of blood pressure.
If the device is required to treat several diseases or condi ons, proceed as fol-
lows: Bring the 1st disease treatment course to comple on, then make a break for
10 to 15 days and
fi
nally proceed to the 2nd disease treatment course. If the device
is required to treat the single disease or condi on, keep the pause of 6 to 8 weeks
between treatment courses. Shorter pauses between treatment courses are only
allowed upon the physiotherapist’s advice.