OPERATION MANUAL
Treatment Modes
The desired treatment mode is set with «Mode Select» bu on as indicated in
Table 1.
Table 1
Opera ng
mode
Field
type
Inductor coil
excita on
frequency
Peak (amplitude)
value of the mag-
ne c
fi
eld density
on the inductor coil
working surface
(mT)
Mode
descrip on
1
traveling
6.25
(1/8 of the
mains
frequency)
20±6
Basic mode
of opera on
2
traveling
6.25
(1/8 of the
mains
frequency)
8±2
Pediatric mode
with reduced
magne c
fi
eld
density
3
sta onary
100
(doubled mains
frequency)
6±2
Therapeu c mode
with pronounced
analgesic and an -
in
fl
ammatory e
ff
ects
The desired mode is selected by pressing the «Mode Select» bu on (Mode 1 –
Mode 2 – Mode 3 – Mode 1) in sequence, and a sound signal is heard repeatedly
where the number of repe ons corresponds to the mode number. Mode indica-
on is provided by corresponding LED indicator.