5 Maintenance
5.2
Cleaning procedure
Cleaning procedure for soiled impeller screw tapped hole:
1. Remove stub shaft (7) per section 4 of Service manual.
2. Submerge and soak the stub shaft for 5 minutes in a COP tank with 2% caustic wash
3. Scrub the blind tapped impeller screw hole vigorously by plunging a clean 1/2” diameter sanitary bristle pipe brush in and out
of the hole for two minutes while submerged.
4. Soak stub shaft (7) in acid sanitiser for 5 minutes, then scrub blind tapped hole as described in step 3 above.
5. Rinse well with clean water and blow-dry the blind tapped hole with clean air.
6. Swab test the inside of the tapped hole to determine cleanliness.
7. Should the swab test fail, repeat steps 2 to 6 above until swab test is passed.
Should swab testing continue to fail, or time is of the essence, install a new (spare) stub shaft (7).