background image

Installation Manual 5850-023

TE91K022-EN3

 

BADK 082012 

Note1:

After 60 seconds without any action from the user the instrument 
switches back to normal mode indicating the actual measured values.

Note 2:

Each of the four measuring ranges is assigned its own setting for 
temperature compensation.

The compensation is adjustable in the range 0.00...4.98%/K.

The compensation is defined as being linear at 25°C.

Note 3: 

Selectable temperature ranges: 
0...150, -20....130, 0...100, -20...80, 0...50, -10...40, -20...150°C

Calibration of the Conductivity Sensor

This procedure describes an easy procedure for determination of the 
temperature coefficient (TC) of a media.
1.  Dip the measuring head in a sample media making sure that the 
channel bore is completely covered with media without bubbles. 
2.  Heat the media to 25.0°C
3.  Ensure adequate circulation in the media.
4.  Note the indicated conductivity.
5.  Heat the media to min. 60°C
6.  Set the TC in the setup menu to indicate exactly the same 
conductivity as it is at 25°C.
Warnings:
1.  A higher TC value will result in a lower conductivity indication.
2.  Do not use the TC value to adjust a measured value. The 
instrument has been calibrated from the factory.
3.  Inadequate circulation in the media will cause a slight heating of the 
channel bore resulting in a false measurement. 

How to contact Alfa Laval Tank Equipment A/S

For further information please feel free to contact: 

Alfa Laval Tank Equipment A/S
Baldershoej 19
P.O. Box 1149
2635 Ishoej
Denmark

Phone no.: +45 43 55 86 00
Fax no.:   +45 43 55 86 01
www.alfalaval.com

Contact details for all countries are continually updated on our 
websites.

Note: The illustrations and specifications contained in this manual 
were effective at the date of printing. However, as continuous 
improvements are our policy, we reserve the right to alter or modify 
any unit specification on any product without prior notice or any 
obligation.

Reservation and contact infomation

© 2012 Alfa Laval Corporate AB 
This document and its contents is owned by Alfa Laval 
Corporate AB and protected by laws governing intellectual 
property and thereto related rights. It is the responsibility 
of the user of this document to comply with all applicable 
intellectual property laws. Without limiting any rights related 
to this document, no part of this document may be copied, 
reproduced or transmitted in any form or by any means 
(electronic, mechanical, photocopying, recording, or otherwise), 
or for any purpose, without the expressed permission of Alfa 
Laval Corporate AB. Alfa Laval Corporate AB will enforce its 
rights related to this document to the fullest extent of the law, 
including the seeking of criminal prosecution.

Service/Repair

Return

Upon every return of the device, no matter if for recalibration, decalcifi

-

cation, modifications or repair, it is necessary to contact your local Alfa 

Laval office to guarantee a quick execution of your request.

Please inform us by sending an email to: 
[email protected].

Include the number of devices sent and request a Return Number. 
Afterwards clean the device, pack it shatterproof and send it to Alfa 
Laval Tank Equipment A/S indicating the Return Number.

Warranty Conditions

The warranty conditions are subject to the legal warranty period of 12 

months from the date of delivery. In case of improper use, modifica

-

tions of or damages to the device, we do not accept warranty claims. 
Damaged diaphragms will also not be accepted. Furthermore, defects 
due to normal wear are not subject to warranty services.

Summary of Contents for TE67K Series

Page 1: ...Alfa Laval Conductivity Transmitter TE67Kxxxxxxxxx have according to the type been developed for applications where conduc tivity and temperature are to be measured Inductive conductivity measurement is based on the principle of two series connected toroidal core transformers 1 and 2 The primary side of the transformer 1 is controlled by an AC voltage generator The liquid flowing through the chann...

Page 2: ... Insert length L 40 Standard sensor shaft Insert length L 87 Extended sensor shaft Mechanical Installation Dimensional Drawing After Installation and Configuration Check the leak tightness of the sleeve Check the tightness of glands or M12 plugs Check the tightness of the cover ø89 Clamp 1 1 2 ISO2852 DIN32676 TEK67K1311111xx TE67K1211111xx Clamp 2 ISO2852 DIN32676 TEK67K1711111xx TE67K1811111xx G...

Page 3: ...our measuring ranges All four measuring ranges can individually be set to a range according to the table The 3 A mark is valid only when the product is mounted in a 3 A marked counter part and installed according to the installation manual Use also a 3 A marked O ring or gasket if relevant The 3 A marked products conforms to the 3 A Sanitary Standard criteria Materials and surfaces fulfill the FDA...

Page 4: ...y Sensor Configuring the Alfa Laval Conductivity Sensor Jog shuttle Step Subject Range Display Turn Push Note 1 Power on starting 2 Actual value 1 yyyyy mS zzzzz C Push 3 SW version Actual version Push 4 Actual value 1 yyyyy mS zzzzz C Right 5 Set Range 1 1 4 20 mA Actual range mS Push Note 1 6 1 4 20 mA Set range mS Left Right Push to set Left Decrease Right Increase 7 1 4 20 mA New range mS Righ...

Page 5: ...ns contained in this manual were effective at the date of printing However as continuous improvements are our policy we reserve the right to alter or modify any unit specification on any product without prior notice or any obligation Reservation and contact infomation 2012 Alfa Laval Corporate AB This document and its contents is owned by Alfa Laval Corporate AB and protected by laws governing int...

Page 6: ...lenght 37 mm TE67Kx7xxxxxxx Clamp DN 50 ISO2852 Clamp DN 51 DIN32676 with sensor shaft lenght 84 mm TE67Kx8xxxxxxx DN 32 Din11851 L 37mm TE67KxAxxxxxxx DN 40 Din11851 L 37mm TE67KxBxxxxxxx DN 50 Din11851 L 37mm TE67KxCxxxxxxx DN 32 Din11851 L 84mm TE67KxDxxxxxxx DN 40 Din11851 L 84mm TE67KxExxxxxxx DN 50 Din11851 L 84mm TE67KxFxxxxxxx Surface finish wetted parts Ra 0 8µ TE67Kxx1xxxxxx Ra 0 4µ Elec...

Page 7: ...UGH December 31 2012 Timothy R Rugh Executive Director 3 A Sanitary Standards Inc The issuance of this authorization for the use of the 3 A Symbol is based upon the voluntary certification by the applicant for it that the equipment listed above complies fully with the 3 A Sanitary Standards designated Legal responsibility for compliance is solely that of the holder of this Certificate of Authoriza...

Page 8: ...Installation Manual 5850 023 TE91K022 EN3 BADK 082012 ...

Reviews: