- Warnings and Error Codes - 

Warnings and Precautions 

 

WAIT AT LEAST 20 MINUTES AFTER EATING OR DRINKING BEFORE TESTING.

 A breathalyzer 

examines “lung air” samples, not “mouth air” to calculate alcohol content. Any “contaminants” in the 
mouth, including alcohol, food, smoke, or even excess saliva, may interfere with testing. This 
includes alcohol-based mouthwash or similar products. 

This is the most common cause of improper 

use and incorrect results.

 

 

Avoid testing in high wind or enclosed spaces. Poor ventilation may also lengthen the time required 
for warm up between tests.

 

 

Do not blow smoke, saliva or other contaminants into the mouthpiece, as the sensor may become 
damaged.

 

 

Do NOT use the Premium as a tool to drink and drive.

 

 

Keep out of extreme temperatures and out of the reach of children.

 

 

To maintain accuracy, the sensor inside the Premium must be recalibrated every 200 tests or at least 
once a year (whichever comes first). Large numbers of "high positive" (high alcohol) readings may 
require recalibration of the sensor much sooner than normally expected.

 

 

Troubleshooting 

 

WARNING DISPLAY

 “

Flow

”: If you see “Flow” in the display, the breath sample was not strong or 

long enough for measurement. Please try again and make sure the breath is strong and steady 
during sampling.  As soon as you start blowing into the device, you will hear a long beep indicating 
how long you should blow.

 

 

WARNING DISPLAY

 “

Cal

”: If you see “Cal” in the display, the Premium needs a new sensor module. 

If 

the cumulative test counter reaches 300 tests, the Premium will display Cal

 

automatically. Please 

purchase and install a new sensor module

.

 

 

WARNING DISPLAY

 “

bAt

”: If you see “bAt” in the display, the battery power is low. Please replace 

the batteries with a new pair.  Make sure to properly align the batteries by looking at the +/- marks.

 

 

WARNING DISPLAY

 "

Err2

":

 If either of these codes appear in the display, the device is reporting a 

general problem with the alcohol sensor, or that the device has been physically damaged.  
Please replace sensors or contact AK GlobalTech Corp for technical support. 

 

WARNING DISPLAY

 "

Err9

"

: Power source is too weak to complete operation.  Please replace 

batteries with a new pair. 

 

For the other troubles, please contact AK GlobalTech at 

[email protected]

, along with your 

error codes and photos if available.   
 

 

Summary of Contents for premium AL-7000

Page 1: ...1 Premium AL7000 AlcoMate Premium AL7000 Operator s Manual ...

Page 2: ...2 ...

Page 3: ...nology With the unique easy replace sensor panel module replacement takes just a few seconds The AlcoMate Premium s PRISMTM Sensor Technology uses advanced next generation components to improve upon prior oxide semiconductor technologies and introduces a level of accuracy reliability and ease of use never before achieved among portable screening breathalyzers Accessories for the BASIC PACK include...

Page 4: ... hold the Power Button down The display may alternately show nor which is short for normal operation The back lit digital display shows results clearly during day or night conditions Release the Power Switch There will be a countdown as the unit warms up The display will show blow when ready Secure your lips around the mouthpiece and blow sharply and steadily into the unit for about 4 seconds If y...

Page 5: ... show is 0 40 A true BAC of 0 40 or above may result in the onset of coma or even death by respiratory arrest A reading of 0 40 may indicate an acutely dangerous condition but may also indicate a sampling error such as liquid alcohol blown into the device Please carefully read the Warning section of this manual before operating the device General Information about DUI DWI The legal limit for intox...

Page 6: ...expected Troubleshooting WARNING DISPLAY Flow If you see Flow in the display the breath sample was not strong or long enough for measurement Please try again and make sure the breath is strong and steady during sampling As soon as you start blowing into the device you will hear a long beep indicating how long you should blow WARNING DISPLAY Cal If you see Cal in the display the Premium needs a new...

Page 7: ...excluding spent or used sensor modules under normal use for one 1 year from the date of purchase AK GlobalTech s obligation under this warranty is limited to replacing adjusting or repairing the unit only if returned along with proof of purchase complete shipping address daytime contact phone number and written explanation The warranty is void of the unit has been tampered with excluding the openi...

Page 8: ...d by service performed by anyone who is not an official representative reseller or distributor of AK GlobalTech or d to a product or part that has been modified to alter functionality or capability without the written permission of AK GlobalTech TO THE EXTENT PERMITTED BY LAW THIS WARRANTY AND THE REMEDIES SET FORTH ABOVE ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES REMEDIES AND CONDITIONS WH...

Page 9: ...This warranty gives you specific legal rights and you may also have other rights that vary by country stat or province This Limited Warranty is governed by and construed under the laws of the country in which the product purchase took place Technical Support AK GlobalTech Corporation 21 Grand Ave Suite 102 Palisades Park NJ 07650 800 943 1016 Toll Free 9A 6P EST 201 943 9411 Office 9A 6P EST 201 9...

Reviews: