Alcatel TA 101 Instruction Manual Download Page 2

1. 

G E N E R A L I T E S

Ces manometres a vide du type 
thermocouple sont composes d'un coffret 
alimentation et d'une jauge. Ils sont 
destines a effectuer des mesures dans la 
gamme des pressions 10' a 1 mbar. 
Leurs caracteristiques essentielles sont 
definies par un code de designation :

1. COMMENTS

Thermocouple-type gauges are composed 
of  a  power  supply  and  a  gauge.  They  are 
designed for making measurements in the 
10

-3

 to 1 mbar range.

Their basic characteristics are defined by 
a designation code:

1. ALLGEMEINES

Das Gerat arbeitet nach dem Princip des 
Thermoelements. Es besteht aus einem 
Netzgerat und der MeBrohre. Der 
MeBbereich reicht von 10

-3

 bis 1 mbar. 

Die wesentlichen Daten gehen aus der 
Typenbezeichnung hervor:

2. CARACTERISTIQUES 
2.1 Alimentation

Secteur 115/220 V, 50/60 Hz commutable 
sur face arriere par deplacement du 
commutateur S2. Cable secteur 
independant de ('alimentation.
L'appareil se presente sous la forme d'un 
coffret 1/4 rack 19" 3U, destine soit a etre 
place dans un chassis rack 19", soit a etre 
encastre dans un panneau, soit a etre 
pose sur table; dans ce cas, it dolt etre 
equipe des pieds caoutchouc livres avec 
chaque appareil.
■ Protection: un fusible calibre place en 
face arriere, valeur: 0,05 A retarde pour 
220 V et 0,1 A retarde pour 115 V.

■ Consommation max secteur: 8 VA

■ Entrée jauge sur la face arriere; cable 
independant de ('alimentation.
■ Sortie  mesure  (enregistrement)  par 
douille 0 2 sur la face arriere: 0 a 10 V 
- 2,5 mA max.
■ Precision  ±  10  %  de  la  lecture  de5.10

-3

a 5.10

-1

 mbar apres etalonnage en air sec 

a 25°C.
■ Une variation de la tension secteur de 
± 10 % entraine une erreur de ± 10 % sur ('
indication de pression dans la zone 10

-2

 a 2.

10

-1

 mbar, plus importante aux

extremites de gamme.
■ Constante de temps de reponse :
- de 1 a 10

-3

 mbar : 30 s,

- de 10

-3

 a 1 mbar : 1 s.

■ Stability globale en temperature de 5 

45°C: 1% / °C 

sur la lecture de pression 

de 5.10

-3

 a 5.10

-1

 mbar.

Poids: TA 101: 1,950 kg

TA 111: 2,000 kg

Sortie seuil

La sortie seuil sur TA 111 est prevue par 
embase J40 sur face arriere. La fiche 
correspondante est fournie. Contact 
inverseur: 100W, 220V, 5A max. (sur circuit 
resistif).

A Parr& de ('alimentation, ce contact est 
sur mauvais vide.
Lorsque la jauge est debranchee ou toupee I'
asservissement indique mauvais vide.

Un voyant rouge s'allume lorsque la 
pression est superieure au point de 
consigne; un voyant vert s'allume dans le 
cas contraire.

2. CHARACTERISTICS 
2.1 Power supply

Check for correct voltage. Line voltage 
setting (115/220V, 50/60Hz) can be 
changed by shifting selector S2 on the 
rear panel. Power cord is independent of 
power supply.
The  unit  is  in  the  form  of  a  1/4  19"  3U 
module  for  installation  in.a  19"  rack  or 
panel  or  on  a  table;  in  the  latter  case  it 
must be fitted with the rubber feet
provided with each unit.
■ Protection: a calibrated fuse located on 
rear panel with a value of: 0,05A (slow-
blow) for 220 V and 0,1 A slowblow for 
115 V.

Maximum power draw : 8 VA.

Gauge input on rear panel; separate 

cable from power supply.

Measurement output for recording 

through 0 2-mm socket on rear panel: 
0 to 10V, 2,5 mA max.
■ Accuracy ± 10 % of reading from 5.10

-3

to 5.10

-1

 mbar after dry-air calibration at 

25°C.

A line-voltage variation of ± 10 % 

causes an error of ± 10 % in the pressure 
reading in the 10

-2

 to 2.10

.1

 mbar zone, 

greater at both ends of the range.

Response time constant :

- 1 to 10

-3

 mbar : 30 sec,

- 10

-3

 to 1 mbar : 1 sec.

Overall temperature stability from 5 to 

45° C: 1%/°C on pressure reading from 
5.10

-3

 and 5.10

.1

 mbar.

Weight: TA 101 - 1.950 kg (4 Ibs) 

TA 111 - 2.000 kg (4.5 Ibs)

Control switch output

On the TA 111, the control switch output 
is J40 socket on the rear panel. The 
corresponding plug is supplied. Transfer 
contact 100W, 220V, 5A max (on 
resistance circuit).

When the power is off, this contact is on "
poor vacuum".
When the gauge is unplugged or 
disconnected, the control reads "poor 
vacuum".

A red light comes on when the pressure is 
above the set point; a green light comes 
on if it is below the set point.

2. TECHNISCHE DATEN 
2.1 Stromversorgung

Die Umschaltung der Netzspannung von 
115/220V, 50/60Hz erfolgt durch 
Umstecken ein Haupschalter S2 auf der 
Ruckseite. Das Netzkabel ist
unabhangig vom Gerat. Das MeBgerat ist 
als 1/4 Rackeinschub 19" 3U ausgefiihrt, 
der entweder in einen 19" Rackrahmen 
geschoben, oder in eine Schalttafel 
eingelassen, oder auf einen Tisch gestellt 
werden kann, in letzteren Falle ist er mit 
den angelieferten GummiffiBen zu 
versehen.

Schutzart: kalibrierte Sicherung auf 

Ruckseite 0,05 A fur 220 V und 0,1 A fur 
115 V, verzogert.

Maximal Verbrauch am Netz: 8 VA.

Mel3rohreneingang auf Ruckseite; Kabel 

unabhangig von der Versorgung.

Registrierausgang fiber Buchsen 2 0 

auf der Gerateruckseite.
AnschluBwerte 0-10 V, 2,5mA max.

MeBgenauigkeit ± 10 % der Anzeige 

zwichen 5.10

-3

 und 5.10

.1

 mbar nach 

Eichung mit trockener Luft bei 25° C.

Netzspannungsschwankungen von 

± 10 % verursachen MeBwert-
schwankungen von ± 10 % zwischen 10

-2

to 2.10

.1

 mbar, an den MeBbereichenden 

sind die Schwankungen groBer.

Ansprechzeit Konstante zwischen :

- 1 und 10

-3

 mbar : 30 sek.,

-10

-3

 und 1 mbar : 1 sek.

Temperaturstabilitat  zwischen  5  und 

45°  C:  1%  der  Anzeige  zwischen  5.10

-3

und 5.10

.1

 mbar.

Gewicht: TA 101 - 1,950 kg

TA 111 - 2,000 kg

Sollwertausgang

Der Sollwertausgang liegt auf einer J40 
Steckerbuchse an der Geraterfickseite. 
Der Umschaltkontakt hat eine 
Schaltleistung von: 100W, 220V, 5A max (
Wiederstandkreis).

Bei abgeschaltetem Gerfit sowie bei 
fehlender Verbindung zur Mellriihre steht 
das Schaltrelais auf der Schaltstellung "
schlechtes Vakuum".

Eine  rote  Lampe  brennt  wenn  der  Druck 
fiber  der  Schaltschwelle  liegt,  eine  grfine 
wenn er darunter liegt.

2