АМM-1130
38
Range and Accuracy
The below-listed accuracies under different ranges refer to those which are guaranteed by the meter within one-year calibration,
with normal use under the operating temperature of 18
℃
-28
℃
and relative humidity less than 80%.The presentation for
accuracy is: ± (**% reading number of lower digits)
●
AC Voltage
Range
Resolution
Accuracy
40Hz-1kHz
220mV
0.01mV
±(0.5% + 20)
2.2V
0.1mV
±(0.8% + 40)
22V
1mV
±(0.8% + 40)
220V
10mV
±(0.8% + 40)
1000V
100mV
±(0.8% + 40)
Notes:
a)In case of probe hanging in the air, the voltage inducted by the testing line may cause unstable readings
on the display screen, but that will not affect the accuracy of measurement.
b)The meter is a true rms responding meter. When the input leads are shorted together in the ac
functions, the Meter may display a residual reading between 1 and 30 counts. It will not attect stated
accuracy, but if using REL to offset this reading may produce a much larger constant error in later
measurements.
Notes: above accuracies can be guaranteed within 10%-100% of the full range.
Summary of Contents for ??M-1130
Page 1: ...AMM 1130 USER S MANUAL www tmatlantic com...
Page 18: ...M 1130 18 Figure 3 2 AC Measurement Warning Not try to measure a voltage higher than 1000 Volt...
Page 20: ...M 1130 20 Figure 3 3 DC AC mV Measurement...
Page 22: ...M 1130 22 Figure 3 4 Logic Frequency Duty Ratio Measurement...
Page 24: ...M 1130 24 Figure 3 5 Resistance continuity Test Diode Measurement...
Page 26: ...M 1130 26 Figure 3 6 Capacitance Measurement...
Page 28: ...M 1130 28 Figure 3 8 DC AC A Measurement...
Page 30: ...M 1130 30 Figure 3 9 DC AC mA Measurement...