4. 

If the resistance is < 50

,

 

a tone will sound. 

Diode Test 

1. 

Connect 

 

as described for resistance measurements 

2. 

Press the 

MODE

 button to select diode test

 

 .

 

3. 

Touch the test probe tips to the diode or semiconductor junction under test. Note the meter reading. 

Reverse the test lead polarity by reversing the red and black leads. Note this reading.The diode or junction can be evaluated as 
follows:If one reading displays a value (typically 0.400V to 01.800V) and the other reading displays 

OL

, the diode is good. If both 

readings display 

OL

 the device is open.If both readings are very small or ‘0’, the device is shorted. 

Capacitance Measurements 

WARNING: 

To avoid electric shock, discharge the capacitor before measuring.  

1. 

Press the 

MODE

 button to select capacitance measurements. 

2. 

Insert the black test lead banana plug into the negative 

COM

 jack. 

Insert the red test lead banana plug into the positive 

 jack. 

3. 

Touch the black test probe tip to one side of the device.  
Touch the red test probe tip to the other side of the device. 

4. 

Read the capacitance value in the display. 

Note: For very large values of capacitance measurement time can be several seconds before the final reading stabilizes. 

Note: The Zero feature removes stray test lead capacitance to improve the accuracy of low value capacitance measurements. To 

perform a zero, Press and hold the 

MODE ZERO

 button for two beeps. The display will zero. The offset value is now stored and is 

removed from all measurements. 

.

Frequency and Duty Ratio Measurements 

1. 

Rotate the function switch to the 

Hz %

 Position. 

2. 

Insert the black test lead banana plug into the negative 

COM

 jack. 

Insert the red test lead banana plug into the positive 

Hz

 jack. 

3. 

Touch the black test probe tip to one side of the device.  
Touch the red test probe tip to the other side of the device. 

4. 

Read the Frequency value on the upper large display. 
Read the Duty Ratio on the lower small display. 

5. 

Press the 

MODE

 button to display the Duty Ratio on the large  

display. 

Power and Measurement 

1. Rotate the function switch to the 

KW/KVA

 Position. 

2. The connecting method see right figure. 

3. Press the 

MODE

 button to select AC or DC power. 

4. Press 

COS

θ

button to select  KVA ,KW,or Power Factor. 

Type K Temperature Measurements  

1. 

Rotate the function switch to the 

Temp

 

position. 

2. 

Press the MODE button to select °F or °C. 

3. 

Plug the transfer socket into the negative COM jack and the positive V jack,  then insert t
he TYPE K into the transfer socket, make sure that the positive and negative  feet are   c
onnected correctly.  

4. 

Connect the temperature probe tip(s) to the object which will be tested.  

5. 

Read the temperature on the display.  

Note: 

In case of an open input  or a temperature overrange, the meter will display “

- - - - -

” .

 

 
Data Hold 

To freeze the LCD reading, press the 

HOLD

 button. While data hold is active, the 

HOLD

 icon appears on the LCD. Press the

 HOLD

 

button again to return to normal operation. 

MAX/MIN  

1. Press the 

MAX/MIN

 button to activate the MAX/MIN recording mode. The display icon "

MAX

" will appear. The meter will begins 

recording and displaying the maximum value measured. 

2. Press the 

MAX/MIN

 button and “

MIN

” will appear. The meter will display the minimum value measured during the recording session. 

3. Press the 

MAX/MIN

 button and “

MAX

 

MIN

” will appear. The meter will display the present reading, but will continue to update and 

store the max and min readings. 

4. To exit MAX/MIN mode press and hold the 

MAX/MIN

 button for 2 seconds. 

 

Summary of Contents for ACM-2352

Page 1: ...Clamp Watt Meter ACM 2352 USER S MANUAL...

Page 2: ...equipment SAFETY NOTES Do not exceed the maximum allowable input range of any function Do not apply voltage to meter when resistance function is selected Set the function switch OFF when the meter is...

Page 3: ...button 5 MAX MIN button 6 INRUSH button 7 Backlit LCD Display 8 Non Contact Voltage Detector 9 NCV LED indicator 10 Function switch 11 KVA KW COS select button 12 Back light button 13 ZERO button 14...

Page 4: ...rent measurements To perform a zero select ADC and with no conductor in the jaw press the ZERO button The display will show zero The offset value is now stored and removed from all measurements Freque...

Page 5: ...egative COM jack Insert the red test lead banana plug into the positive Hz jack 3 Touch the black test probe tip to one side of the device Touch the red test probe tip to the other side of the device...

Page 6: ...o the OFF position and then turn to the desired function position To disable APO 1 From the OFF position hold the MODE button and rotate the FUNCTION switch to a measurement function 2 will appear in...

Page 7: ...True RMS 50 Hz to 1000 Hz 400 0 mVAC 0 8 9digits 50 60Hz 4 0000 VAC 1 0 30 digits 40 000 VAC 400 00 VAC 750 0 VAC All AC voltage ranges are specified from 5 of range to 100 of range Resistance 400 00...

Page 8: ...ating Humidity Max 80 up to 31 C 87 F decreasing linearly to 50 at 40 C 104 F Storage Humidity 80 Operating Altitude 7000ft 2000meters maximum Battery One 1 9V Battery NEDA 1604 Auto power OFF After a...

Reviews: