275g001e_SC5010_HS_ES_3rd ed.docx
2022-01-11
Page 5 of 22
2
Useful life and warranty conditions
The useful product life is expected to be 8 years. Both, the useful life time and the warranty offered
by manufacturer are conditional upon the adherence to the conditions outlined below.
3
Scope of delivery
The SC 5010 HS / ES scope of delivery comprises
Position
Qty
SC 5010 HS / ES examination chair
1
Re charger unit incl. power cord
1
Battery
1
Quality- test report
1
User Manual
1
Options according to order
4
Intended use
The medical device is intended for use as a surgical chair. Any other use of the product is not
permitted.
o
The medical device is designed for a maximum patient weight of 200 kg (440 lbs) (EN 60601-
2-46 particular requirements for safety of operating tables). A 300 kg (660 lbs) version is
available upon request.
o
The headrest is designed for a maximum load of 20 kg (44 lbs).
o
The chair is suitable to be employed in the immediate proximity of patients.
Caution
Risk of
operating
errors
The device may only be installed, operated, used and maintained by
persons who have been properly trained or who have the required
knowledge and experience to do so. Please also adhere to the
national qualification guidelines applicable in your country.
Caution
General
Hazards
The development, production and maintenance of these medical
devices, together with associated risks, are based on an expected
service life of eight years, provided the device is serviced at the
specified intervals.
Modifications to the product or failure to follow the manufacturer’s
instructions may substantially reduce the expected service life and
significantly increase the risks associated with the use of this device.
It is the responsibility of the institution operating this product to
follow the manufacturer’s instructions and to decide on the
risk/benefit ratio upon expiration of the expected service life or
maintenance and inspection intervals specified by the
manufacturer.
Caution
General
Hazards
The product is only operated in compliance with the provisions of
the intended use and the environmental conditions set forth in the
chapter “Ambient conditions”