297G001e.docx
2019-04-15
Page 6 of 10
5
Mounting and commissioning
The medical device is delivered completely assembled. If optional parts or accessories are part of the
delivery, please refer to the respective section in this user manual for mounting or connecting in-
structions.
If not expressively mentioned in this user manual, no additional assembly or calibration work is re-
quired to operate the medical device.
6
Electrical connections
The location of the power input of the device or the recharging unit is described in this user manual.
To connect the medical device to the power supply use only the power cable included in this deliv-
ery.
Warning
General Haz-
ard
The medical device is only de-energized when the mains plug has
been pulled.
The device should be set up so that the power cable can be discon-
nected from the power supply quickly and easily without any tools.
Warning
Risk of elec-
tric shock
Additional extension cables or portable multiple sockets may not be
used.
The electrical installation must conform to IEC 60364-7-710 or the
applicable national regulations. This includes the existence of a
ground fault circuit interrupter (GFCI).
To avert the risk of an electric shock this device may only be connect-
ed to a power supply with a protective earth conductor.
Ensure that the power supply plug is suitable and certified for the
local connection. If the supplied power cable must be replaced, at
least the following specifications must be adhered to:
• Protective ground conductor resistance of the power cable maxi-
mum 0.1 Ohm
• Local certification of the power cable for connection to medical
devices
• Device plug C19 conforming to IEC 60320
• Cross
-section at least 1.5 mm2/AWG 16
Warning
Fire Hazard
The device is not suitable for operation in explosion risk areas (e.g.
combustible mixture of anesthetic, cleaning or disinfecting agents
with air, oxygen or nitrous oxide).
The electrical installation must conform to IEC 60364-7-710.
The data concerning power consumption on the type plate must be
taken into account in the selection of over current protection.
7
Mounting
After mounting connect powercord with the chair. The connector is under the baseplate. Put plug
into socket
Caution
Don’t roll over the powercord. Look after the cord. Use only cords in
good condition. Don’t move chair without disconnecting.
Summary of Contents for ak 5003M
Page 12: ......
Page 13: ...Mammography Chair ak 5003M User Manual...
Page 23: ...Poltrona per visita mammografica Mammography Chair ak 5003M Manuale d uso User Manual...
Page 34: ......
Page 35: ...Silla de mamograf a Mammography Chair ak 5003M Manual del operador User Manual...
Page 46: ......
Page 47: ...Fauteuil pour mammographie Mammography Chair ak 5003M Mode d emploi User Manual...