Gas Pre-Alarm activated. Flashing together with the gas concentration (See page 12).
Setup wizard initiated (See page 9).
Gas alarm activated. Flashing together with the gas concentration (See page 12).
Error or malfunction of the sensor. Please contact your technical support centre.
Request for protection code (See pages 21 and 22).
SA reading of greater than 300 PPM has been detected during the reset to zero
process (See page 21).
The temperature of the sensor has experienced a variation of greater than 5 °C during
the calibration or reset to zero process (See pages 21 and 22).
The maximum time has been exceeded without the reading stabilising during the
calibration process (15 min) or reset to zero process (5 min) (See pages 21 and 22).
The calibration or reset to zero process has been completed successfully (See pages 21
and 22).
The sensor has reached its maximum working temperature.
Calibrating or resetting to zero (See pages 21 and 22).
The sensor has reached its minimum working temperature.
Initialisation process of the gas sensor. This process may last for up to 3 minutes.