![AirSep Centrox Instruction Manual Download Page 48](http://html1.mh-extra.com/html/airsep/centrox/centrox_instruction-manual_2879051048.webp)
48
Centrox Instruction Manual
AirSep Corporation
8.1
TROUBLESHOOTING CHART
The chart on the following pages is a guide for troubleshooting the AirSep Centrox
Oxygen Concentrator.
The interior of the oxygen concentrator contains electrical parts
that can produce an electrical hazard if not handled properly. To
prevent electrical shock, use extreme care when you service the
system.
The Printed Circuit Boards (PCBs) contain components that are
sensitive to electrostatic discharge (ESD) and can be damaged if
not handled properly. As when handling any ESD-sensitive PCB,
observe standard ESD safety procedures. These procedures
include the following:
■ Handle the PCB only by the edges
■ Work on a grounded ESD mat
■ Wear a grounded wrist strap
■ Store PCBs only in anti-static bag
Problem
Probable Cause
Solution
Oxygen compressor(s) do
not start. No green power
light.
PSA module not plugged to
the power supply.
Plug in the PSA module.
Compressor module not
plugged in to the PSA
module.
Plug in compressor module
to the PSA module.
Tripped Circuit breaker(s).
Reset circuit breaker(s).
ON/OFF switches on the
compressor module are in
OFF position.
Put the ON/OFF switches on
the compressor module in
ON position.
ON/OFF switch on the PSA
module in OFF position.
Put the ON/OFF switch
on the PSA module in ON
position.
Defective ON/OFF switch.
Replace ON/OFF switch.
Blown 1A or 15A fuse in PSA
module.
Replace the appropriate
fuse.
Faulty electrical connections.
Check electrical connections.
Defective circuit board.
Replace circuit board. (Refer
to Section 8.2.2).