![AirSep Centrox Instruction Manual Download Page 29](http://html1.mh-extra.com/html/airsep/centrox/centrox_instruction-manual_2879051029.webp)
Centrox Instruction Manual
29
AirSep Corporation
5.2
PRE-INSTALLATION GUIDELINES
Before you install the Centrox oxygen concentrator, and the oxygen receiver, if
supplied, refer to the Specifications section in the Appendix of this instruction
manual to determine the applicable space, and the power requirements for your
particular model.
A backup source of oxygen must be available if a power failure or
system malfunction occurs.
Make sure the area that surrounds the oxygen concentrator is
well ventilated, and provide sufficient space around the unit [at
least three feet] to allow for cool air flow as well as to allow safe
operation and maintenance.
Locate the oxygen concentrator in an area where the ambient
air temperature remains between 4°C (40°F) and 40°C (104°F) to
prevent damage not covered under the AirSep Corporation Product
Warranty.
Connect the oxygen concentrator power cord only to a properly
grounded electrical outlet on a circuit that cannot be accidentally
turned off. Do not use extension cords.
Provide proper voltage to the oxygen concentrator to prevent
damage not covered under the AirSep Product Warranty.
Do not plug in the power cord until you complete the installation of
the oxygen concentrator.