![AirSep Centrox Instruction Manual Download Page 24](http://html1.mh-extra.com/html/airsep/centrox/centrox_instruction-manual_2879051024.webp)
24
Centrox Instruction Manual
AirSep Corporation
Kit #
Description
Qty
.
KI046-1
Primary/Secondary Ball Valve Assembly
1
Hose Assembly, Oxygen-clean, (Secondary Hose)
1
KI470-1
Regulator-Flowmeter Assembly
1
Hose Assembly, Oxygen-clean (Main Oxygen Outlet Hose)
1
TA150-1
60 Gallon Tank Assembly, O
2
Cleaned
1
Table 4.2:
Description of the Parts included in the Starter Kits
Please contact an AirSep Corporation Sales representative for ordering the
starter kits. Below is a brief description of items included in the starter kits.
Primary Oxygen Ball Valve
The primary oxygen ball valve controls the flow of oxygen from the oxygen
concentrator to your oxygen distribution system. See Figure 5.1.
Secondary Oxygen Ball Valve
The secondary oxygen ball valve controls the flow of a backup source of oxygen to
your oxygen distribution system. This ball valve should always be closed unless a
backup source is connected to it. See Figure 5.1.
Regulator-Flowmeter Assembly
This assembly consists of a pressure regulator and a flowmeter. The pressure
regulator attaches to the outlet port of the oxygen receiver to regulate the pressure
of the oxygen. The flowmeter is used to regulate the flow of the oxygen at the
outlet of the oxygen receiver. Refer to Figure 5.1 for the installation location of this
assembly.
Oxygen Isolation Ball Valve Assembly
The oxygen isolation ball valve stops the flow of oxygen to the oxygen receiver
during troubleshooting. The oxygen relief valve in the assembly prevents excess
pressure from building in the oxygen receiver if a system malfunction occurs. This
assembly is shipped mounted on the oxygen receiver. See Figure 4.5.
Main Oxygen Hose
An oxygen-clean hose is provided that connects the outlet of the supply valve
assembly (Refer to section 4.5) on the oxygen concentrator to the oxygen isolation
ball valve assembly on the oxygen receiver.
Secondary Oxygen Hose
Secondary hose is provided to connect the outlet of the regulator-flowmeter
assembly to the inlet of the primary oxygen ball valve. Refer to Figure 5.1.