Pressure relief position

In pressure relief position of particular body parts, in operating mode ‘Static’, cells can be
removed individually. To do this, you have to remove the cover of the mattress, pull out
the push buttons from the cells to be removed, left and right from the base mattress.
Tear the connection hose of the cell, to the hose system in the mattress base loose. You
now close, with the exit cover (article number 10-8007) the resulting opening at the hose
system, in order to avoid a leakage including its alarm. Pull, the temporarily cells pulled
off, from the cell regulation and connect the cover to the mattress (base system).

At a pressure relief position, in operating mode ‚Alternating pressure‘, double as large,

not filled, free areas may arise, in case the adjacent cell for non-pressure position remo-

ved cell, for the duration of the interval cycle is not filled. In this case, the patient bed

sores.

Security key

Symbol for active security key in the display:

The security key will activate automatically after ca. 5 minutes. You will recognize the

active security key by the symbol, as shown in the display. Pressing a key, in case the of

an activated security key, please follow the instructions shown in the display.

You will deactivate the security key by pressing button

(9).

If the security key in the

display is not visible, you can adjust the settings. The security key is activating itself,

after ca. 5 minutes after the last touch of the key or by pressing key

(9)

.

The security key function is to prevent unintended operation of the device. Turning off

the device is not prohibited by the security key

(2)

. Nevertheless, the user should press

key

(2)

 for 2 seconds consciously, to turn off the device

Set the interval time

Setting the interval time is only possible in operating mode ‘Alternating pressure’. An

interval time from 3 till 30 minutes can be set, each time with a step of one minute. The

interval time is needed to switch cell to cell.

The desired interval time is set, by pressing key

(12)

. For a shorter interval period press

'-' for a longer interval period press '+'. The period of time interval is displayed in minu-

tes on the display.

Set the weight of the patient

Setting the weight of the patient is both possible in operating mode ‚Alternating pressure‘

and ‚Static‘.

The weight of the patient is set by key

(10)

. The steps are based on entire kilograms.

For a lower weight of the patient press '-' and for a higher weight of the patient, press

'+'.

Summary of Contents for ADS clinic

Page 1: ...Bedienungsanleitung Mode d emploi Wechseldrucksystem Syst me anti escarre ADS Clinic...

Page 2: ...ce 12 Choice of operating mode 14 Nursey mode 14 Pressure relief position 15 Security key 15 Set the interval time 15 Set the weight of the patient 15 Display of operating hours 16 Turning off the dev...

Page 3: ...the ADS clinic is linked to a special type of matress from AirMedPLUS The ADS clinic is microprocessor controlled device equip ped with an active pressure control and keeps the pressure in the air ce...

Page 4: ...page 4 of 22 Version 1 4 June 2015 ADS_Clinic Operator Manual _V1_4_Mai 2015_Software 1_20_ENG doc Product images Front view 1 2 3 4 5 6 7 8 9 10 11 12 Rear view 14 13...

Page 5: ...page 5 of 22 Version 1 4 June 2015 ADS_Clinic Operator Manual _V1_4_Mai 2015_Software 1_20_ENG doc Top view 15 16 17 18 19 16 14 20 21 13 22 Display 23 24 25 26 23 27 25...

Page 6: ...page 6 of 22 Version 1 4 June 2015 ADS_Clinic Operator Manual _V1_4_Mai 2015_Software 1_20_ENG doc 28 29 Patient s mattress 30 31 32 33 34 36 37 35 38 39...

Page 7: ...Powercord 15 Screw 16 Bed suspension 17 Vibration damper 18 Pump shock absorbers 19 Type plate 20 Fuse 1 A 250 V 21 Fuse 1 A 250 V 22 Dustfilter Display 23 View of the selected mattress type 24 Shows...

Page 8: ...l should be available and undamaged You can order the O rings at the service center as a replacement part item number 99 1131 Make sure all hoses are at one direction of the pneumatic plug and whether...

Page 9: ...stance is limited The top straps may only the be connected at the placeable part of the mattress It s not allowed to connect the mattress and the bed using the top straps Otherwise there is danger tha...

Page 10: ...es in the area or ambient air Please note The device should not be covered There is a danger in case the device is covered The heat will accumulate Please note Alarm function will only work when ADS c...

Page 11: ...ng the device briefly hanging on the foot of bed only and exclusively to operate the device and not during the use On behalf of security reasons it should be established that no humidity can run insid...

Page 12: ...max 6 cm foam mattress foam not included REF 10 8215 20 Kg 180 Kg 2 Mattress replacement system support stop in June 2015 50 Kg 210 Kg 3 Mattress replacement system REF 10 8335 70 Kg 250 Kg You chang...

Page 13: ...e static and two filled rooms and one non filled room if you have selected operating mode alternating pressure If the symbol shown in the upper left corner of the display no longer flashes the filling...

Page 14: ...3 30 min set in steps of one minute the cell pattern A B C A B C will for the duration of the set interval time inflate the cells A and B C not After the interval time cell A is deflated cell B remai...

Page 15: ...ty key by the symbol as shown in the display Pressing a key in case the of an activated security key please follow the instructions shown in the display You will deactivate the security key by pressin...

Page 16: ...device To turn of the device press for 2 seconds key 2 Key 2 can also be pressed in case the key security is switched on The power off function will take precedence over the security key The alarm Th...

Page 17: ...don t use the device anymore Please contact our service department Please note Alarm function will only work when ADS clinic is conneted to the electricity supply system If the electricity supply sys...

Page 18: ...tions The profile of the patient s mattress is changed in case the AD system is in alternating mode For certain patient s the mobilization can lead to health problems In order to avoid contraindicatio...

Page 19: ...ares herewith that the following medical device Product name ADS clinic Meets the requirements of the European directive for medical products 93 42 EWG and 2007 47 EC The ADS clinic is consistent with...

Page 20: ...page 20 of 22 Version 1 4 June 2015 ADS_Clinic Operator Manual _V1_4_Mai 2015_Software 1_20_ENG doc Information manufacturer and technical details...

Page 21: ...ient 40 Kg 13 2 mmHg Maximum weight patient 210 Kg 42 7 mmHg pressure range 13 2 42 7 mmHG maximum fill pressure 42 7 mmHg Number of cells 21 length of the cells ca 90 cm width of the cells ca 9 5 cm...

Page 22: ...e und Dienstleistungen f r Heime Spit ler und Spitex Produits r ha et services pour homes h pitaux et organismes de soins domicile S gestrasse 75 CH 3098 K niz Telefon 0848 10 20 40 Fax 031 970 0171 i...

Reviews: