Choice of operating mode

The ADS

clinic

has two operating modes to select i.e.:

“Static mode” and “Alternating mode”

Static mode

The symbol identification in the display of the mode:

By pressing button (3) the function “Static mode” of the ADS clinic is selected and the

operating will start. In this function, the ADS clinic works like a therapeutically static

function. All cells are filled with the same selected pressure, via the set weight of the

patient.

In static therapeutic function and the carefully selected weight of the patient, the patient

descends deeper in the mattress. By increasing the surface, the pressure load is spread

over multiple planes, and for the individual the risky areas of the body reduced.

Alternating mode

The symbol identification in the display of the mode:

By operating, key

(4)

, the function of the ADS

clinic

 changes into a three room alterna-

ting pressure system. By a user-set interval time (3-30 min., set in steps of one minute)

the cell pattern A-B-C-A-B-C.. will, for the duration of the set interval time, inflate the

cells A and B, C not. After the interval time, cell A is deflated, cell B remains filled, and

cell C will be inflated. At the end of the next interval time, cell B will be inflated, cell C

remains fill and cell A is refilled. The filling/deflating of cells causes 'air flow sound‘, by

the so-called ‚Airshifting‘ procedure. At ' Airshifting ' is the volume moved from one room

to another room. So has the ADS

clinic

 on short interval times for a momentary peak

load.

Nursey mode

By  simultaneous  pressing  of  buttons

(8)

  and

(9)

  for  approximately  three  sconds  you  will

enable or disable the "nursery mode". "Nursery mode" means, all cells will be inflated with

maximum pressure for  maximum 20 minutes. In the display, you will see the icons „Rest“

und „MAX“ (flashing), see picture:

After  20  minutes  in  "nursery  mode",  „ADS  clinic“  will  turn

automatically  back  to  the  previous  mode  before  enabled  the

"nursery  mode".  The  remaining  time  is  shown  into  the  display

with the word „Rest“. If nursery mode is enabled, you can turn

into the previous mode at any time by pressing the buttons

(8)

and

(9)

simultaneously for approximately 3 seconds.

Summary of Contents for ADS clinic

Page 1: ...Bedienungsanleitung Mode d emploi Wechseldrucksystem Syst me anti escarre ADS Clinic...

Page 2: ...ce 12 Choice of operating mode 14 Nursey mode 14 Pressure relief position 15 Security key 15 Set the interval time 15 Set the weight of the patient 15 Display of operating hours 16 Turning off the dev...

Page 3: ...the ADS clinic is linked to a special type of matress from AirMedPLUS The ADS clinic is microprocessor controlled device equip ped with an active pressure control and keeps the pressure in the air ce...

Page 4: ...page 4 of 22 Version 1 4 June 2015 ADS_Clinic Operator Manual _V1_4_Mai 2015_Software 1_20_ENG doc Product images Front view 1 2 3 4 5 6 7 8 9 10 11 12 Rear view 14 13...

Page 5: ...page 5 of 22 Version 1 4 June 2015 ADS_Clinic Operator Manual _V1_4_Mai 2015_Software 1_20_ENG doc Top view 15 16 17 18 19 16 14 20 21 13 22 Display 23 24 25 26 23 27 25...

Page 6: ...page 6 of 22 Version 1 4 June 2015 ADS_Clinic Operator Manual _V1_4_Mai 2015_Software 1_20_ENG doc 28 29 Patient s mattress 30 31 32 33 34 36 37 35 38 39...

Page 7: ...Powercord 15 Screw 16 Bed suspension 17 Vibration damper 18 Pump shock absorbers 19 Type plate 20 Fuse 1 A 250 V 21 Fuse 1 A 250 V 22 Dustfilter Display 23 View of the selected mattress type 24 Shows...

Page 8: ...l should be available and undamaged You can order the O rings at the service center as a replacement part item number 99 1131 Make sure all hoses are at one direction of the pneumatic plug and whether...

Page 9: ...stance is limited The top straps may only the be connected at the placeable part of the mattress It s not allowed to connect the mattress and the bed using the top straps Otherwise there is danger tha...

Page 10: ...es in the area or ambient air Please note The device should not be covered There is a danger in case the device is covered The heat will accumulate Please note Alarm function will only work when ADS c...

Page 11: ...ng the device briefly hanging on the foot of bed only and exclusively to operate the device and not during the use On behalf of security reasons it should be established that no humidity can run insid...

Page 12: ...max 6 cm foam mattress foam not included REF 10 8215 20 Kg 180 Kg 2 Mattress replacement system support stop in June 2015 50 Kg 210 Kg 3 Mattress replacement system REF 10 8335 70 Kg 250 Kg You chang...

Page 13: ...e static and two filled rooms and one non filled room if you have selected operating mode alternating pressure If the symbol shown in the upper left corner of the display no longer flashes the filling...

Page 14: ...3 30 min set in steps of one minute the cell pattern A B C A B C will for the duration of the set interval time inflate the cells A and B C not After the interval time cell A is deflated cell B remai...

Page 15: ...ty key by the symbol as shown in the display Pressing a key in case the of an activated security key please follow the instructions shown in the display You will deactivate the security key by pressin...

Page 16: ...device To turn of the device press for 2 seconds key 2 Key 2 can also be pressed in case the key security is switched on The power off function will take precedence over the security key The alarm Th...

Page 17: ...don t use the device anymore Please contact our service department Please note Alarm function will only work when ADS clinic is conneted to the electricity supply system If the electricity supply sys...

Page 18: ...tions The profile of the patient s mattress is changed in case the AD system is in alternating mode For certain patient s the mobilization can lead to health problems In order to avoid contraindicatio...

Page 19: ...ares herewith that the following medical device Product name ADS clinic Meets the requirements of the European directive for medical products 93 42 EWG and 2007 47 EC The ADS clinic is consistent with...

Page 20: ...page 20 of 22 Version 1 4 June 2015 ADS_Clinic Operator Manual _V1_4_Mai 2015_Software 1_20_ENG doc Information manufacturer and technical details...

Page 21: ...ient 40 Kg 13 2 mmHg Maximum weight patient 210 Kg 42 7 mmHg pressure range 13 2 42 7 mmHG maximum fill pressure 42 7 mmHg Number of cells 21 length of the cells ca 90 cm width of the cells ca 9 5 cm...

Page 22: ...e und Dienstleistungen f r Heime Spit ler und Spitex Produits r ha et services pour homes h pitaux et organismes de soins domicile S gestrasse 75 CH 3098 K niz Telefon 0848 10 20 40 Fax 031 970 0171 i...

Reviews: