Connecting a Transducer

All transducers must be connected to the TDT1000 through a test cable

. If the transducer has a 

connector, you need a purchased OEM test cable. It must match the brand of the instrument and the 
number of pins in the connector of the transducer.
If the transducer terminates in bare wires, go to “Transducer with NO Connector.”

Transducer with Connector

Test Cable

IMPORTANT

: Each OEM has specific test cables. Be sure you have the correct test cable. Many 

cables look alike or have the same connector.

Always use the SensorCheck app to determine the correct test cable.

1. Open the SensorCheck app.
2. Select the three horizontal bars in the upper left corner (Figure 4).

3. From the drop-down menu, select “Search Part Number.” This will bring up the search field.
4. Input the Airmar part number located on the transducer tag. As you enter this number the app 

will display the correct test “Cable to use” part number for that transducer.

Figure 4. Determining the correct test cable

Copyright© 2020 Airmar Technology Corporation

transducer
part number

test cable
part number

11

Summary of Contents for TDT1000

Page 1: ...TDT1000 D 17 615 01 rev 4 17 615 01 rev 4 09 23 20 Transducer Diagnostic Tester Owner s Guide...

Page 2: ...tronic or printed copy of this manual or revision must contain the complete text of this copyright notice and provided further that any unauthorized commercial distribution of this manual or any revis...

Page 3: ...5 Introduction 5 Kit Includes 6 Test Cables 7 Where to Purchase Cables 7 Test Set up 8 TDT1000 Set up 10 Connecting a Transducer 11 About Airmar s SensorCheck App 15 Troubleshooting 16 Replacement Pa...

Page 4: ...IMPORTANT Before removing and or returning an Out of Range transducer please contact your service representative for additional product support 4...

Page 5: ...s fire hazard if damaged defective or improperly used Exposure to liquids especially saltwater can cause internal corrosion damage to the cells or to the battery management system Batteries should be...

Page 6: ...rope Test Cable with breakout box for transducers with NO connector Test Block Carrying Case Owner s guides The detailed Owner s Guide can be found at www airmar com tdt1000 The SensorCheck app guide...

Page 7: ...B LR connector 33 1333 01 Lowrance Conventional Test Cable 7FB LR connector 33 1332 01 Mix Match Test Cable 1kW 9F A connector 33 1323 01 Mix Match Test Cable 600W 5F A connector 33 1389 01 Mix Match...

Page 8: ...y become permanently deformed Always remove the transducer clean the Test Block and store it in a safe place Good Test Using the TDT Test Block in Air The Test Block simulates 1 to 2m 3 to 6 of water...

Page 9: ...the Test Block by 2 5cm 1 3 Carefully place the acoustic face of the transducer onto the Test Block 4 Gently move the transducer from side to side to fully wet its acoustic face Be careful not to agi...

Page 10: ...ector on the charging cable into the power adapter Figure 3 2 Plug the power adapter into a power supply 3 Plug the micro USB connector on the charging cable into the USB port on the TDT1000 4 When th...

Page 11: ...have the correct test cable Many cables look alike or have the same connector Always use the SensorCheck app to determine the correct test cable 1 Open the SensorCheck app 2 Select the three horizonta...

Page 12: ...there are some exceptions that are not evident when using the transducer Best practice is to connect a Mix Match transducer using the dedicated Mix Match TDT1000 Test Cable rather than connecting the...

Page 13: ...ect it 3 Plug the transducer connector into the test cable Figure 7 4 Plug the test cable into the test cable port on the TDT1000 Twist and lock the connector Figure 6 Test cable with yellow tag Copyr...

Page 14: ...esponding terminals in the breakout box If the test results do not look correct try swapping the solid and striped pairs and testing again Transducers built for OEMs can have different wire color codi...

Page 15: ...ith the test set up or the instrument Do NOT remove return the transducer Contact your customer service representative To summarize information about a transducer and the vessel in which it is install...

Page 16: ...ith breakout box for transducers with NO connector 33 736 01 Carrying Case ACC TDT1000 CASE Quick Start Guide 17 632 01 Owner s Guide can be found at www airmar com tdt1000 17 615 01 SensorCheck app g...

Page 17: ...35 Meadowbrook Drive Milford New Hampshire 03055 4613 USA www airmar com...

Reviews: