14.4 Calculation table
If the measured values deviate from the standard values, the values are specified in chapter "4 Tech-
nical data"
The safety distances can then be calculated in the tables shown below
P:
V
1
:
E
1
:
P
Rated power of the transmitter in Watts (W) in accordance with the specifications of the
transmitter manufacturer
V
1
Compliance level for the test according to IEC 61000-4-6
E
1
Compliance level for the test according to IEC 61000-4-3
Interference im-
munity tests
IEC 60601 – test
level
Compliance level
Recommended safety dis-
tances
Conducted HF dis-
turbance variables in
accordance with
IEC 61000-4-6
3 V
eff
150 kHz to 80 MHz
[V
1
] V
d = [35 / V
1
]
⋅ √
P
Emitted HF distur-
bance variables in
accordance with
IEC 61000-4-3
3 V/m
80 MHz to 25 GHz
[E
1
] V/m
d = [35 / E
1
]
⋅ √
P
for 80 MHz to 800 MHz
d = [7 / E
1
]
⋅ √
P
for 800 MHz to 25 GHz
Rated power of the
transmitter (W)
Safety distance depending on the transmission frequency (m)
150 kHz to 80 MHz
d = [3.5 / V
1
]
⋅√
P
80 MHz to 800 MHz
d = [3.5 / E
1
]⋅√
P
800 MHz to 2.5 GHz
d = [7 / E
1
]
⋅√
P
001
01
1
10
100