
ELECTRA-flo
Air Monitor Corporation
116-053-00.P65 (7/06)
ELECTRA-flo
- IO&M Manual
5.12 – TRANSMITTER OUTPUT CALIBRATION (con't)
6. Once an acceptable zero reading is obtained, press and then and display will indicate:
Output 1 Span
Selection
7. Press . Display will indicate:
Output 1 Span
Perform Calibration
8. With DMM still connected as in Step 4, reading should be at maximum value; 10.00 ± 0.01 volts, 5.00 ± 0.01
volts, or 20.00 ± 0.1mA.
9. If DMM is reading out of tolerance, use or to adjust the ELECTRA-flo's output for an acceptable DMM
reading.
10. Once an acceptable span reading is obtained, press and then and display will indicate:
Output 2 Zero
Selection
11. Repeat Steps 3 through 10 for Transmitter Output 2. For Transmitter Output 2, connect DMM across terminals 3
and 4.
12. Press and display indicates:
Return to
MAIN MENU
21
SECTION 5 – OPERATION
ENT
UP
DN
UP
ENT
UP
Þ
ENT
UP
Þ