![Air Liquide Soffio Cube Instruction Manual Download Page 12](http://html1.mh-extra.com/html/air-liquide/soffio-cube/soffio-cube_instruction-manual_2875011012.webp)
1. Fill the nebuliser with the inhalation solution as per your doctor’s instructions (Fig.B). Ensure
that you do not exceed the maximum level.
2. Assemble the nebuliser and connect the mouthpiece or the nosepiece or the mask. (Fig.C).
Ensure that all parts are complete.
3. Connect the nebuliser with the connecting tube (14) to the compressor (1) and plug the power
cable (2) into the socket.
4. To turn on the device, push the ON/OFF switch (3) onto the position «I». Lean the mouthpiece
between the lips or the nosepiece on the nostrils or lean the mask to the face over mouth
and nose.
5. Sit in a relaxed position with the upper body upright. Breathe in and out calmly during the
therapy. Do not lie down while inha¬ling. Stop inhalation if you feel unwell.
6. After completing the inhalation period recommended by your doctor, switch the ON/OFF switch
(3) to position «O» to turn off the device and unplug it from the socket.
7. Empty the remaining medication from the nebuliser and clean the device as described in the
section «Cleaning, disinfecting and sterilizing».
3 - CLEANING, DISINFECTING AND STERILIZING
Always wash your hands well before cleaning and disinfecting the accessories.
Do not expose the compressor to water or heat.
Replace the connecting tube for each treatment with a new patient or in case of impurities.
Unplug the device before cleaning.
Cleaning
The compressor unit (1) and the connecting tube (14) should be cleaned with a clean, moist cloth.
Clean all the nebuliser components and its accessories under warm tap water (max. 60°C)
adding if necessary a small quantity of detergent following dosage and use limitations as
provided by detergent manufacturer. Rinse thoroughly making sure that all deposits are washed
away and leave to dry.
Disinfecting
The compressor unit (1) can be disinfected with a clean cloth and some chemical disinfectant
following dosage and use limitations as provided by disinfectant manufacturer.
All the accessories (except the connecting tube) can be disinfected with chemical disinfectants
following dosage and use limitations as provided by disinfectant manufacturer. Disinfectants are
usually available at pharmacies.
Steam sterilizing
The nebuliser and its accessories (except air hose and masks) can be heat steam sterilized
up to 121 °C (20 min.) or 134 °C (7 min.). EN554/ISO11134.The sterilization packaging must
conform to EN868/ISO11607 and be suitable for steam sterilization. After sterilization always let
all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when components are still warm.
4 - MAINTENANCE
The SOFFIO CUBE aerosol therapy device does not require any particular maintenance. The
compressor does not need to be lubricated as the piston is the dry-operating type. It is advisable
replace nebulizer after some 100 to 120 treatments on single patient or after about 20 sterilization
cycles. Check the filter continually for cleanliness and replace it if dirty, or after a maximum of 3
months use. Spare filters are provided with the device. To replace the filter, open the air filter slot
(4) (if needed using a tool) and replace the filter (Fig. D).
ENGLISH • ENGLISH • ENGLISH • ENGLISH • ENGLISH • ENGLISH • ENGLISH • ENGLISH • ENGLISH • ENGLISH • ENGL
12