background image

enter the +/- % that the tool will be tested 
by. Press 

Save

 after entering values. 

 

Units 

Various engineering units, cycle from 
INOZ, INLB, FTLB, Nm, CNm, gfcm, Kgfcm 
and kgfm. The check box ensures that as 
engineering units are changed, the Target 
and Limits are also updated. 
 

Mode Setting 

By selecting the proper 

Operating Mode

 

the user can properly take torque 
readings from a variety of manual and 
powered tools.  

 

Peak  

This mode will provide a display of 
the maximum torque value achieved 
by the tool during operation. This 
mode is used for all continuous drive 
tools and click-type torque wrenches.  

 

1

st

 Peak  

This mode will detect and display the 
“first peak” achieved by click wrench-
es and cam-over screwdrivers.  

 

Pulse  

This mode will display the maxi-
mum torque value achieved by 
discontinuous drive tools such as 
pulse tools and impacting tools. 

 

 

Track 

This mode will display torque in 
real-time as it is applied to the 
connected transducer. Track mode 
is used primarily for calibration of 
the unit or identifying the condition 
of the transducer. In Track Mode 
with no torque on the transducer, 
the display should show zero. 
 

Sign Lock 

Setting  allows the user to select the 
rotation direction in which a torque will be 

captured. With Sign Lock 

ON

, the initial 

direction is set as the default direction; a 
negative torque value indicates a reading 
taken in the counter-clockwise direction 
and a positive reading indicates a reading 
taken in the clockwise direction. Any 
readings taken in the non-selected 
direction will be measured, but not 
captured / displayed as a peak value. 
 
With Sign Lock turned 

OFF

, peak values in 

both the clockwise and counter-clockwise 
directions will be captured and displayed. 
Since most torque rundowns are 
conducted and readings are taken in the 
clockwise direction, the user may find it 
beneficial to turn this feature 

ON

 

Frequency 

Sets Frequency filter setting the analyzer 
uses in monitoring the transducer signal. 
For testing Continuous drive power tools, 
500hz is recommended. For testing 
Discontinuous drive tools 1000Hz is 
recommended.  Consult the  specific  tool 
manufacturer for any other tool types. 
 

Peak Blank 

Determines the minimum threshold at 
which a torque peak is  captured. It is 
entered as a percentage of the full scale 
value of the transducer which is 
connected to the ATDA-8000 and can be 
displayed in values from 2 –  50% of 
transducer full scale.  
 

Dimming 

Sets display timer 
 
After choosing the settings of the unit select 
the 

Done

 button and the unit will return to 

the run mode with enabled settings. You 
can also Restart or Turn Off the unit.

 

 

Summary of Contents for AUDITOR ATDA-8000 Series

Page 1: ... ATDA 8000 Series Torque Data Analyzers User Guide ...

Page 2: ...unctions 7 8 Operating Instructions Basic Navigation 9 Transducers Transducers and Torque Measurement Cable 11 12 Rundown Kits 12 Software Auditor Tool Manager 13 14 Auditor Data Manager 15 Unit Specifications Dimensions Width 7 Height 5 Depth 1 Weight 1 Lb Power Requirements Main Power 100 240VAC 50 60hz from supplied charger Operating Temperature Range 0o to 50o C Data Communication Serial ...

Page 3: ...o servicing Lock Out Tag Out procedures should be followed To avoid electric shock which may result in personal injury and or death cords should be kept from heat sharp edges or any other potentially hazardous conditions When operating a power tool outdoors use an extension cord suitable for outdoor use Keep out of reach of children DANGER Contact with water solvents or other liquid substances may...

Page 4: ...st Run Button 7 Battery Indicator 8 Preset Tool Dropdown Main Screen illustrates Target Sample Reading Mode Engineering Unit Full Scale value of Transducer s Sample Auto Clear setting Peak Blanking setting Test Results Auto Clear setting Low Limit setting Sign lock setting High Limit setting Frequency Response setting Number of Samples Captured and Battery Indicator 3 6 4 5 2 1 7 8 ...

Page 5: ... Setting 2 Results Auto Clear Setting 8 Sign Lock 3 Samples Per Set 9 Frequency Setting 4 Threshold Setting 10 Peak Blank Setting 5 6 Sigma Adjustment 11 Display Dimming Timer 6 Engineering Units Setting 12 Turn Off Done Restart Buttons 2 3 1 7 4 8 9 6 11 5 12 10 ...

Page 6: ...6 Figure 3 Results Screen Results Screen illustrates Target Torque Value Test PASS FAIL High Low Average Standard Deviation and Six Sigma percentage Recorded samples and indicates Pass Fail ...

Page 7: ...e sample set or auto save complete sample set If left Off the user will be required to manually clear the readings by pressing the Clear Test Run button at the bottom of the display in the Main screen Samples per set Sets number of samples to be taken per audit Choose how many torque samples per test by pressing the or buttons You may choose a maximum of 18 Target Sets Target Nominal torque settin...

Page 8: ... which a torque will be captured With Sign Lock ON the initial direction is set as the default direction a negative torque value indicates a reading taken in the counter clockwise direction and a positive reading indicates a reading taken in the clockwise direction Any readings taken in the non selected direction will be measured but not captured displayed as a peak value With Sign Lock turned OFF...

Page 9: ...upply port 2 Plug the Charger into a 100 240 VAC power outlet using the appropriate plug end 3 Prior to turning unit on connect the transducer to an Auditor supplied cable ICBL 8000DIG via any open USB port on the ATDA 8000 4 To turn the ATDA 8000 on off press the ON OFF button Figure 1 5 The Charging Mode Indicator will illuminate as you are charging and show status 6 The ATDA must be used with t...

Page 10: ...ed After a test run is completed the test data can be cleared manually or by using the Results Auto Clear timer to automatically clear results Each of the following buttons can be used to execute a command Select Auto or Manual mode Auto Mode allows user to o Clear readings o Zero the display transducer o Enter readings into Samples Captured screen Manual Mode allows user to o Select torque mode o...

Page 11: ...the port Tighten the retaining bars to prevent cable from disengaging from transducer or instrument Power Cable End Transducer Cable End Setting up the Transducer The Auditor Torque Data Analyzer will recognize the transducer and report the full scale in the display Most transducers can be made to allow the Auditor Torque Data Analyzer to self recognize the transducer simply when a connection is m...

Page 12: ...e Model Description In Lb Nm Drive IN ARDIA 10 HD HDE HDS Rundown Fixture 1 0 10 13 1 13 1 4 ARDIA 25 HD HDE HDS Rundown Fixture 2 5 25 28 2 8 1 4 ARDIA 100 HD HDE HDS Rundown Fixture 10 0 100 1 3 11 3 1 4 ARDIA 250 HD HDE HDS Rundown Fixture 25 0 250 2 8 28 25 3 8 ARDIA 500 HD HDE HDS Rundown Fixture 50 0 500 5 6 56 5 3 8 ARDFA 100 HD HDE HDS Rundown Fixture 10 100 13 6 136 1 2 ARDFA 150 HD HDE H...

Page 13: ...eate a database of tools and store test results with this easy to use software integrator to from tester to PC Tool Set up is easy with touchscreen access to type Tool Name Serial Asset Capacity Model and Manufacturer information Type of Tool and Capacity Engineering Units have selections from drop down boxes when triangle is touched Once details are entered touch Save Tool button to save data ...

Page 14: ...re are a wide range of additional very specialized functions that the ATDA 8000 series systems can accommodate Hydraulic tools measurement of pressure observed through a pressure transducer and more can be audited with these products Specific needs and training for these higher level functions capabilities require specialized hardware Please consult your authorized AIMCO Auditor representative to ...

Page 15: ...ta Click Audit Manager to open Connect USB data stick to any open USB slot on the ATDA 8000 series analyzer Look to bottom right corner of the ATDA 8000 screen USB symbol should show to the left of the battery condition symbol indicating the USB stick has successfully mounted to the device Select desired data to be exported by either o Selecting specific audit o Selecting date range o Selecting al...

Page 16: ...ORATION DE MEXICO SA DE CV 10000 SE Pine Street Ave Cristobal Colon 14529 Portland Oregon 97216 Chihuahua Chihuahua 31125 Phone 503 254 6600 Mexico Toll Free 1 800 852 1368 Phone 01 614 380 1010 LIT MAN873 Rev 07 2020 Printed in USA 2020 AIMCO ...

Reviews: