15
Auditor™ Data Manager Software
Auditor Data Manager Software is included with the purchase ATDA-8000 series torque
analyzers. This software is imbedded into the unit. The View Data tab allows you to
look at the collected data. This data can be exported to Excel as a .csv file and
displayed by Excel charts. When selecting a test run - the average, range, sigma, Cp,
and CpK values are listed.
To export data:
•
Click Audit Manager to open.
•
Connect USB data stick to any open USB slot on the ATDA-8000 series analyzer.
•
Look to bottom right corner of the ATDA-8000 screen. USB symbol should show to
the left of the battery condition symbol indicating the USB stick has successfully
mounted to the device.
•
Select desired data to be exported by either
o
Selecting specific audit
o
Selecting date range
o
Selecting all data
•
Press Export Selected Data.
•
Dialog box will display indicating successful transfer of data to the connected USB
stick
•
USB stick may now be removed and taken to a PC for download and archiving.