71
CALIBRATION <cpd> [
,
nrf] The calibration control command. <cpd> can be one of three
sub
−
commands:
−
START
Enter calibration mode; this command must be issued before any
other calibration commands will be recognised.
SAVE
Finish calibration, save the new values and exit calibration mode.
ABORT
Finish calibration, do not save the new values and exit calibration
mode.
<nrf> represents the calibration password. The password is only
required with CALIBRATION START and then only if a non
−
zero
password has been set from the instrument’s keyboard. The
password will be ignored, and will give no errors, at all other times.
It is not possible to set or change the password using remote
commands.
CALADJ <nrf>
Adjust the selected calibration value by <nrf>. The value must be in
the range
−
100 to +100. Once an adjustment has been completed
and the new value is as required the CALSTEP command must be
issued for the new value to be accepted.
CALSTEP
Step to the next calibration point.
For general information on remote operation and remote command formats, refer to the following
sections.