AII AII-3000 A Manual Download Page 6

3

 

Safety Warnings

 

 

ALWAYS 

follow the statements below as they are essential to reduc-

ing the risk of use error due to ergonomic features of the device or the 
environment in which the device is intended to be used. 

 

Only  trained  personnel  who  have  read,  understand  and  agree  to  follow 
the Instructions for Use should operate the device.  

 

Retain the Instructions for Use for future reference. 

 

Refer service needs to trained authorized personnel. Failure to do so may 
cause the device to fail and void the warranty. 

 

Inspect  the  device  and  accessories  before  operating  and  ensure:  (a) 
there is no evidence of physical damage; (b) the sensor (particularly the 
sensing surface) and electrical connections are dry; and, (c) the sensor is 

installed and upstream from any humidifying device for accurate calibra-

tion and oxygen readings. 

 

Calibrate:  (a)  with  a  known  source  of  air  or  dry  100%  oxygen  before 
using each day or after 8 hours of continuous use; (b) when the temper-
ature or pressure of the operating environment changes; (c) if the oxy-

gen sensor has been disconnected and reconnected; (d) after the battery 
or oxygen sensor has been replace. 

 

Sampling  flowing  gas:  (a)  install  the  optional  flow  diverter  and  tee-
adapter  in  a  vertical  position  as  shown  in  Section  4.3  and  (b)  assure 

there is a tight fit between the flow diverter and tee adapter. 

 

Sampling  static,  ambient  or  controlled  atmospheres  remove  the  flow 
diverter. 

 

Clean the device and accessories in accordance with Section 6.1.2

 

 

Battery replacement Section 6.2: (a) replace the batteries within twenty-
four (24) hours of the battery symbol appearing on LCD display

 

and (b) 

calibrate the analyzer after replacing the batteries. 

 

Oxygen  sensor  installation  or  replacement  Section  6.3:  allow  the  new 
sensor to stabilize for 15-20 minutes in ambient air before attempting to 
calibrate. 

 

Store the device by turning the power OFF and removing the batteries if 
the device will not be operated for over thirty (30) days. 

 

Attempt  to  repeat  the  procedure  that  caused  a  perceived  malfunction  
and  refer  to  troubleshooting  hints  in  Section  7  before  concluding  the 

device is faulty. If in doubt, contact the manufacturer for assistance. 

NEVER

  operate  the  device  in  any  manner  described  below  doing  so 

may compromise the clinical condition or the safety of patients, users 

or other persons. 

 

If the reading is unstable or a malfunction is suspected. 

 

After the battery symbol appears in the LCD display. 

 

Near equipment capable of emitting high levels of electromagnetic radia-
tion (EMI) or radio frequency interference (RFI). 

 

Expose  the  device;  particularly  the  LCD  display  or  sensor  to  sources  of 

extreme heat, cold or excessive sunlight beyond the device’s storage tem-

perature range, refer to Section 8 for extended periods of time. 

 

In a gas stream with a vacuum greater than 14” water column.

 

 

Immerse the device, oxygen sensor or coiled cable in any liquid. 

 

Outside of the parameters specified in Section 8 particularly at flow rates 
greater than 10 liters per minute - the backpressure generated produces 

erroneously high oxygen readings. 

 

Calibrate:  (a)  with  20.9%  oxygen  or  room  air  with  the  intent  of  taking 
oxygen measurements at oxygen levels above 30% oxygen; (b) in a hu-
midified gas stream or atmosphere; (c) without allowing a newly installed 

sensor to stabilize for 15-20 minutes in ambient air. 

 

Attempt  to  sterilize,  autoclave,  liquid  sterilize,  immerse  in  any  liquid  or 
expose  the  device  or  accessories  to  steam,  ethylene  oxide  or  radiation 

sterilization. 

 

In the presence of flammable gases. 

 

Open the main compartment of the device, except to change the integral 
oxygen sensor of the AII-3000 AHC or AII-3000 MHC Oxygen Analyzers. 

 

Open the oxygen sensor or probe the sensing surface, refer to Section

 

10 

in the event the sensor should leak and someone comes  in contact with 

the electrolyte from inside the sensor. 

 

Operate  with  a  cable  that  appears  worn,  torn  or  cracked,  or,  allow  an 

excess length of cable near the patient’s head or neck; secure it to the bed 

rail or other suitable object to avoid the possibility of strangulation. 

 

Allow the device or oxygen sensor to be serviced, repaired or altered by 
anyone  except  trained  personnel 

  failure  to  do  so  may  endanger  the 

patient or damage the device rendering the warranty null and void. 

姚经理19961810948(微信同号)

QQ:2316517588

邮箱:[email protected]

Summary of Contents for AII-3000 A

Page 1: ...FF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 75 x 1 375 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing 0 95 RH LED Indicators A models None M models upon activation of alarms Linearity 1 under constant conditions Pressure Inlet A M models ambient HC models regulate Vent all models atmos...

Page 2: ...HC AII 11 60 Oxygen Sensor AII 11 60 HC Oxygen Sensor BATT 1008 Battery 2x 1 5V AA BATT 1008 Battery 2x 1 5V AA P 1087 Instructions for Use P 1087 Instructions for Use A 1162 PCB Assy Main A 1162 PCB Assembly Main CABL 1006 Coil Cable TUBE 1019 7 32 OD Tubing 3 CC 1072 Carrying Case O2 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Page 3: ...e mail sales medical aii1 com web www aii1 com This manual may not be reproduced in whole or in part without the prior written consent of Analytical Industries Inc ISO 9001 2008 Certificate 485A AII 3000 AHC AII 3000 MHC AII 3000 M AII 3000 A 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Page 4: ...tanks intended for scuba diving for personnel safety area monitoring O2 deficiency in confined spaces checking breathing air tanks and confirming the O2 levels prior to welding Users must read the following statements as they are essential to re ducing the risk of use error due to ergonomic features of the device or the environment in which the device is intended to be used The devices have been d...

Page 5: ...enting the partial pressure of oxygen in the gas stream being analyzed The resulting oxygen reading is displayed by a large easy to read backlit liquid crystal display LCD that has a resolution of 0 1 oxygen The microprocessor is controlled from a keypad and provides features like system diagnostics warning indicators controls and an alarm capability for continu ous monitoring that enhance both sa...

Page 6: ...cluding the device is faulty If in doubt contact the manufacturer for assistance 4 NEVER operate the device in any manner described below doing so may compromise the clinical condition or the safety of patients users or other persons If the reading is unstable or a malfunction is suspected After the battery symbol appears in the LCD display Near equipment capable of emitting high levels of electro...

Page 7: ...ot to damage the case when removing the battery compartment cover 4 2 1 Procedure 1 Remove the device and the 2 AA 1 5V Alkaline batteries from the foam shipping container 2 Turn the device over so the shortest raised line on the battery compartment cover is pointing away from you 3 Lift the tripod wire stand up and away from the case 4 Grasp the case with both hands use your thumbs press down fir...

Page 8: ...0 AHC MHC When the HC hose connection version is ordered the device is shipped with the sensor installed 8 4 4 Controls 4 4 1 AII 3000 A AHC Oxygen Analyzers These analyzers employ a micro processor that is controlled by five 5 pushbuttons located on the keypad attached to front cover 1 ON OFF provides power to the electronics 2 ESCAPE aborts a previous selected option 3 ENTER selects a menu optio...

Page 9: ... A AHC Oxygen Analyzers The LCD alternately displays the following until the problem is corrected Corrective action 1 Expose the sensor to air or a gas containing approximately 20 9 oxygen 2 Connect or replace the cable connecting the sensor to the analyzer 3 Connect or replace the oxygen sensor 4 5 2 2 AII 3000 M MHC Oxygen Analyzers Performs the same routine and requires the same corrective acti...

Page 10: ...he LO alarm and proceed to the HI alarm or press the UP DOWN arrow keys to change the alarm set point 6 Press ENTER to save LO alarm set point and move to select the HI alarm 7 Press ENTER to skip the HI alarm and return to SAMPLING mode or press the UP DOWN arrow keys to change the alarm set point 8 Press ENTER to save HI alarm set point and return to SAMPLING mode 9 If no key is pressed within 5...

Page 11: ...ixty 60 seconds to allow the sensor to stabilize before the microprocessor takes the final reading 14 7 If the calibration is successful the display below left appears for three 3 seconds before defaulting to the display below right 8 The display above right requires a decision by the user refer to warnings at the beginning of section 4 7 to press ENTER and skip the 100 O2 calibration and return t...

Page 12: ...he sensor itself requires no maintenance and is simply replaced at the end of its useful life like a battery Inasmuch as the sensor is a transducer in its own right its ex pected life is not affected by whether the analyzer is ON or OFF The relationship between the sensor s signal and changes with the oxygen concentration is both proportional and linear thus allowing single point calibra tion Othe...

Page 13: ...he temperature and pressure altitude at which the ana lyzer will be operated 18 Effect of Humidity The analyzer is not affected by non condensing relative humidity RH Howev er the use of a humidifier to introduce water vapor and increase the moisture level of the gas mixture does affect the oxygen concentration and the resultant reading displayed by the analyzer The addition of water vapor increas...

Page 14: ...g displayed by the LCD If placing the entire sensor inside the controlled atmosphere review Section 5 2 Application Consideration Effect of Temperature 5 4 3 AII 3000 AHC and MHC Integral Oxygen Sensor AII 3000 AHC and MHC with their integral oxygen sensor requires connecting the tubing supplied section 4 2 1 above with the device to a hose barb attached to a pressure regulator controlling a sourc...

Page 15: ... simply find and register the male plug at the end of the coiled cable and insert it into the mating female jack at the rear of the sensor until it mates or snaps into place 3 Calibrate the device after replacing the oxygen sensor 6 3 2 Procedure AII 3000 AHC and AII 3000 MHC Integral Sensor 1 Tools required small bladed screwdriver 2 Place the device face down on a flat surface 3 Remove the two 2...

Page 16: ...ction Reading displayed by LCD drifts during calibration Wait 5 minutes and repeat calibration with sensor placed on flat surface not in your hand Check integrity of gas delivery system Check sensor s front o ring seal Verify calibration gas in not humidified Remove moisture covering sensor Replace sensor repeat calibration Analyzer reading climbs after calibration in 100 dry oxy gen when exposed ...

Page 17: ...ration Air or 100 oxygen after 8 hours of continuous use Compensation Temperature compensated Connections A M models 1x16mm thread HC models Tubing 1 4 Controls Soft touch keypad for ON OFF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 5 x 1 5 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing ...

Page 18: ... or in use The purpose of the Recommended Storage Life period is to ensure the user derives the Expected Life of 60 months 915 420 oxygen hours and does not lose the benefit of the warranty Warranty The 12 month 183 084 oxygen hours warranty period begins with ship ment from the factory and is limited to the first claim submitted is based on 8 1 Spare Parts Accessories AII 3000 A AII 3000 M AII 30...

Page 19: ...cidental or consequential losses or damages damages resulting from alterations misuse abuse lack of proper maintenance unau thorized repair or modification of the analyzer Service Contact us between 8 00am and 5 00pm PST Monday thru Thursday or before 12 00pm on Friday Trained technicians will assist you in diagnosing the prob lem and determining the appropriate course of action 31 10 Material Saf...

Reviews: