AII AII-3000 A Manual Download Page 5

In order to obtain optimum performance, the operation of the device must be 
performed in accordance with these Instructions for Use. Maintenance should 

be performed only by trained personnel authorized by the manufacturer. 
 

1.2  Intended Use 

The  AII-3000  Series  Oxygen  Analyzers  are  intended  to  measure  and  display 

the  concentration  of  oxygen  in  compressed  breathing  air  tanks  intended  for 

scuba diving, for personnel safety, area monitoring, O2 deficiency in confined 

spaces,  checking  breathing  air  tanks  and  confirming  the  O2  levels  prior  to 
welding.  

 

1.3  Device Description 

The  AII-3000  Series  Oxygen  Analyzers  can  be  positioned  on  a  table  top  or 
pole (tripod wire stand and V-mount dovetail attachments are mounted on the 

back  of  the  device)  and  are  readily  portable  from  one  location  to  another. 

They provide continuous, fast, reliable and accurate oxygen measurements of 
up to respiratory care systems. 

 
The devices utilize an electrochemical galvanic fuel cell type oxygen sensor of 

the type that is extensively used to measure oxygen concentrations from 0% 
to 100% in gas streams. Oxygen, the fuel for this electrochemical transducer, 

diffusing into the sensor through a gas permeable membrane reacts chemical-
ly at the sensing electrode to produce an electrical current output proportional 

to  the  oxygen  concentration  in  the  gas  phase.  The  sensor  has  an  absolute 
zero  meaning  that  when  no  oxygen  is  present  to  be  chemically  reacted  the 

LCD displays 00.0 oxygen. 
 

The  sensor’s  signal  output  is  linear  over  the  entire  range,  remains  virtually 

constant over the specified useful life and drops off sharply at the end. The 

sensor itself requires no maintenance and is simply replaced at the end of its 
useful  life  like  a  battery.  Inasmuch  as  the  sensor  is  a  transducer  in  its  own 

right, its expected life is not affected by whether the analyzer is ON or OFF. 

 
A battery powered state-of-the-art micro-

processor converts the sensor’s sig-

nal output representing the partial pressure of oxygen in the gas stream being 
analyzed.  The  resulting  oxygen  reading  is  displayed  by  a  large  easy  to  read 

backlit liquid crystal display (LCD) that has a resolution of 0.1% oxygen. The 
microprocessor is controlled from a keypad and provides features like system 

diagnostics, warning indicators, controls  and an alarm capability for continu-
ous monitoring that enhance both safety and effectiveness. 

  
Prior to shipment, every device is thoroughly tested at the factory and docu-

mented  in  the  form  of  a  Quality  Control  Certification  that  is  included  in  the 

Instructions for Use supplied with every device. 

2

 

Quality Control Certification

 

 

Customer: 

________________________  Order No. _____________  Date: _______ 

Model: 

(   )  AII-3000 A  Oxygen Analyzer 

(   )  AII-3000 M  Oxygen Analyzer                     
(   )  AII-3000 AHC  Oxygen Analyzer 
(   )  AII-3000 MHC  Oxygen Analyzer                 S/N _______________ 

Sensor: 

(   )  AII-11-60  or  (   )  AII-11-60-HC                S/N _______________ 

Electronics:  

A-1152  PCB Assembly Main            Software Version _______________ 

Accessories: 

AII-3000 A / M:  CABL-1006  Cable, Coiled Phone Jack         

                                 
AII-3000 AHC / MHC:  TUBE-

1019 7/32” OD Tubing 3 ft.

 

 

All units:  BATT-1008  Battery, 1.5V AA (Qty 2)             
              P-

0187  Manual, Instructions for Use ……… Included  ______

 

 

                                                                                            PASS 

QC Test: 

LCD display 3-

1/2 digits …………………………………………………….   ______

 

 

Battery symbol displays when battery is low ………………………..  ______

 

 

Span adju10-30% FS with 100% oxygen

 

calibration    ______

 

 

Following calibration with 99-100% oxygen and flushing with  
ambient air, oxygen reading as displayed by LCD 20.9% +2%

  ____

 

 

Span adju10-

30% FS with air calibration ………………..  ______

 

 

 

Following calibration with air (20.9% oxygen) and exposing 

to 99-100% oxygen, LCD displays 100% +

2% …………………….  ______

 

 

 

Overall inspection for physical defects ………………………………...  ______

 

 

 

 Qty 

Options: 

_____________________________________________ 

____ 

See Sec 8.1 

____________________________________________________  _____ 

 

____________________________________________________  _____ 

 

____________________________________________________ 

____ 

 

____________________________________________________  _____ 

Delivery: 

____________________________________________________   

姚经理19961810948(微信同号)

QQ:2316517588

邮箱:[email protected]

Summary of Contents for AII-3000 A

Page 1: ...FF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 75 x 1 375 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing 0 95 RH LED Indicators A models None M models upon activation of alarms Linearity 1 under constant conditions Pressure Inlet A M models ambient HC models regulate Vent all models atmos...

Page 2: ...HC AII 11 60 Oxygen Sensor AII 11 60 HC Oxygen Sensor BATT 1008 Battery 2x 1 5V AA BATT 1008 Battery 2x 1 5V AA P 1087 Instructions for Use P 1087 Instructions for Use A 1162 PCB Assy Main A 1162 PCB Assembly Main CABL 1006 Coil Cable TUBE 1019 7 32 OD Tubing 3 CC 1072 Carrying Case O2 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Page 3: ...e mail sales medical aii1 com web www aii1 com This manual may not be reproduced in whole or in part without the prior written consent of Analytical Industries Inc ISO 9001 2008 Certificate 485A AII 3000 AHC AII 3000 MHC AII 3000 M AII 3000 A 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Page 4: ...tanks intended for scuba diving for personnel safety area monitoring O2 deficiency in confined spaces checking breathing air tanks and confirming the O2 levels prior to welding Users must read the following statements as they are essential to re ducing the risk of use error due to ergonomic features of the device or the environment in which the device is intended to be used The devices have been d...

Page 5: ...enting the partial pressure of oxygen in the gas stream being analyzed The resulting oxygen reading is displayed by a large easy to read backlit liquid crystal display LCD that has a resolution of 0 1 oxygen The microprocessor is controlled from a keypad and provides features like system diagnostics warning indicators controls and an alarm capability for continu ous monitoring that enhance both sa...

Page 6: ...cluding the device is faulty If in doubt contact the manufacturer for assistance 4 NEVER operate the device in any manner described below doing so may compromise the clinical condition or the safety of patients users or other persons If the reading is unstable or a malfunction is suspected After the battery symbol appears in the LCD display Near equipment capable of emitting high levels of electro...

Page 7: ...ot to damage the case when removing the battery compartment cover 4 2 1 Procedure 1 Remove the device and the 2 AA 1 5V Alkaline batteries from the foam shipping container 2 Turn the device over so the shortest raised line on the battery compartment cover is pointing away from you 3 Lift the tripod wire stand up and away from the case 4 Grasp the case with both hands use your thumbs press down fir...

Page 8: ...0 AHC MHC When the HC hose connection version is ordered the device is shipped with the sensor installed 8 4 4 Controls 4 4 1 AII 3000 A AHC Oxygen Analyzers These analyzers employ a micro processor that is controlled by five 5 pushbuttons located on the keypad attached to front cover 1 ON OFF provides power to the electronics 2 ESCAPE aborts a previous selected option 3 ENTER selects a menu optio...

Page 9: ... A AHC Oxygen Analyzers The LCD alternately displays the following until the problem is corrected Corrective action 1 Expose the sensor to air or a gas containing approximately 20 9 oxygen 2 Connect or replace the cable connecting the sensor to the analyzer 3 Connect or replace the oxygen sensor 4 5 2 2 AII 3000 M MHC Oxygen Analyzers Performs the same routine and requires the same corrective acti...

Page 10: ...he LO alarm and proceed to the HI alarm or press the UP DOWN arrow keys to change the alarm set point 6 Press ENTER to save LO alarm set point and move to select the HI alarm 7 Press ENTER to skip the HI alarm and return to SAMPLING mode or press the UP DOWN arrow keys to change the alarm set point 8 Press ENTER to save HI alarm set point and return to SAMPLING mode 9 If no key is pressed within 5...

Page 11: ...ixty 60 seconds to allow the sensor to stabilize before the microprocessor takes the final reading 14 7 If the calibration is successful the display below left appears for three 3 seconds before defaulting to the display below right 8 The display above right requires a decision by the user refer to warnings at the beginning of section 4 7 to press ENTER and skip the 100 O2 calibration and return t...

Page 12: ...he sensor itself requires no maintenance and is simply replaced at the end of its useful life like a battery Inasmuch as the sensor is a transducer in its own right its ex pected life is not affected by whether the analyzer is ON or OFF The relationship between the sensor s signal and changes with the oxygen concentration is both proportional and linear thus allowing single point calibra tion Othe...

Page 13: ...he temperature and pressure altitude at which the ana lyzer will be operated 18 Effect of Humidity The analyzer is not affected by non condensing relative humidity RH Howev er the use of a humidifier to introduce water vapor and increase the moisture level of the gas mixture does affect the oxygen concentration and the resultant reading displayed by the analyzer The addition of water vapor increas...

Page 14: ...g displayed by the LCD If placing the entire sensor inside the controlled atmosphere review Section 5 2 Application Consideration Effect of Temperature 5 4 3 AII 3000 AHC and MHC Integral Oxygen Sensor AII 3000 AHC and MHC with their integral oxygen sensor requires connecting the tubing supplied section 4 2 1 above with the device to a hose barb attached to a pressure regulator controlling a sourc...

Page 15: ... simply find and register the male plug at the end of the coiled cable and insert it into the mating female jack at the rear of the sensor until it mates or snaps into place 3 Calibrate the device after replacing the oxygen sensor 6 3 2 Procedure AII 3000 AHC and AII 3000 MHC Integral Sensor 1 Tools required small bladed screwdriver 2 Place the device face down on a flat surface 3 Remove the two 2...

Page 16: ...ction Reading displayed by LCD drifts during calibration Wait 5 minutes and repeat calibration with sensor placed on flat surface not in your hand Check integrity of gas delivery system Check sensor s front o ring seal Verify calibration gas in not humidified Remove moisture covering sensor Replace sensor repeat calibration Analyzer reading climbs after calibration in 100 dry oxy gen when exposed ...

Page 17: ...ration Air or 100 oxygen after 8 hours of continuous use Compensation Temperature compensated Connections A M models 1x16mm thread HC models Tubing 1 4 Controls Soft touch keypad for ON OFF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 5 x 1 5 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing ...

Page 18: ... or in use The purpose of the Recommended Storage Life period is to ensure the user derives the Expected Life of 60 months 915 420 oxygen hours and does not lose the benefit of the warranty Warranty The 12 month 183 084 oxygen hours warranty period begins with ship ment from the factory and is limited to the first claim submitted is based on 8 1 Spare Parts Accessories AII 3000 A AII 3000 M AII 30...

Page 19: ...cidental or consequential losses or damages damages resulting from alterations misuse abuse lack of proper maintenance unau thorized repair or modification of the analyzer Service Contact us between 8 00am and 5 00pm PST Monday thru Thursday or before 12 00pm on Friday Trained technicians will assist you in diagnosing the prob lem and determining the appropriate course of action 31 10 Material Saf...

Reviews: