AII AII-3000 A Manual Download Page 13

5.2  Application Considerations 
 

Effect of Temperature 

All membrane clad electrochemical sensors are temperature dependent due to 

the expansion and contraction of the Teflon sensing membrane. As result more 
or less of the sample gas including oxygen to be reacted diffuses into the sen-

sor.  The  oxygen  sensor’s  electrical  current  signal  output  varies  linearly  with 

oxygen concentration. The signal also varies with changes in ambient tempera-

ture. The temperature coefficient is typically 2.54% of the signal or reading per 
degree C change in temperature. 

 
The temperature dependent current signal output is  compensated by using a 

resistor-thermistor  network.  With  a  proper  resistor-thermistor  network,  the 
signal can be compensated to 5% of the oxygen reading over the 5-

45°C temperature range. This is the worse case situation when going from one 

extreme  of  the  operating  temperature  range  to  the  other.  The  error  will  be 
eliminated when the thermistor in the temperature compensation network and 

the electrolyte inside the sensor reach thermal equilibrium in approximately 45-
60 minutes. 

 

Erroneous  oxygen  readings  can  result  if  the  gases  flowing  over  the 

sensing area of the sensor are not at ambient temperature. This occurs 
because the sensor is exposed to different temperatures. The sensing 

area of the sensor is o-ring sealed in the heated breathing circuit and the tem-
perature compensation network at the rear of the sensor is exposed to ambient 

temperature. 
 

Effect of Pressure 

Electrochemical sensors actually measure the partial pressure, not the percent-

age, of oxygen in the gas stream they are exposed to. These sensors are accu-
rate  at  any  pressure  provided  the  pressure  is  constant  and  the  analyzer  has 

been calibrated at the same pressure as the sample gas measured. 

 
For example, when moving an analyzer calibrated at sea level into the moun-

tains  causes  the  analyzer  to  display  an  decrease  in  the  oxygen  reading  dis-
played. When if fact, the decrease in the reading displayed is not related to a 

change  in  the  oxygen  percentage  but  to  the  decrease  in  partial  pressure 
(corresponding to the increase in total pressure) at altitude. 

 
Calibrate at the temperature and pressure (altitude) at which the ana-

lyzer will be operated. 

18 

Effect of Humidity 

The analyzer is not affected by non-condensing relative humidity (RH). Howev-

er, the use of a humidifier to introduce water vapor and increase the moisture 
level of the gas mixture does affect the oxygen concentration and the resultant 

reading  displayed  by  the  analyzer.  The  addition  of  water  vapor  increases  the 
total  pressure  thereby  diluting  or  decreasing  the  oxygen  concentration  of  the 

gas mixture resulting in a lower oxygen reading.  

 

Effect of Condensation 

Excessive condensation collecting on the sensing area or the electrical connec-

tions at the rear of the sensors can adversely impact the performance of elec-
trochemical sensors. Condensation blocks the diffusion path of oxygen into the 

sensor and can reduce the oxygen reading to 00.0 if the condensation covers 
the entire sensing area. Condensation on the electrical connections at the rear 

of the sensor can affect oxygen readings. Remedy either situation by shaking 

out the condensation and allowing the sensor to air dry. 
 

Erroneously characterized in many instances as a sensor failure, excessive con-
densation is remedied by gently wiping away the condensation with a soft cloth 

or simply allowing the sensor to air dry. 
 

Effect of Electromagnetic Radiation 

Tested over a 26 MHz to 1000 MHz electromagnetic field, the analyzer is sus-

ceptible at all frequencies tested except those between 930 and 990 MHz. 
 

Never  operate  the  analyzer  near  equipment  capable  of  emitting  high 
levels of electromagnetic radiation. Do not continue to operate the ana-

lyzer if the reading becomes unstable. 

 

5.3  Calibration 

Calibrating the analyzer or monitor during normal operation involves the same 

precautions and procedures as those described in Sections 4.7 Start-up Calibra-

tion with the same cautions to review Sections 3 Safety Warnings and 5.2 Ap-
plication Considerations. 

 

5.4  Sampling 

Assuming the START-UP instructions are followed and the tests are completed 
successfully the devices default to the SAMPLING mode. 

 
Never operate the analyzer if the reading is unstable or if a malfunction 

is  suspected. If calibration is  required as indicated herein, do not pro-
ceed until the analyzer is calibration successfully. 

19 

姚经理19961810948(微信同号)

QQ:2316517588

邮箱:[email protected]

Summary of Contents for AII-3000 A

Page 1: ...FF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 75 x 1 375 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing 0 95 RH LED Indicators A models None M models upon activation of alarms Linearity 1 under constant conditions Pressure Inlet A M models ambient HC models regulate Vent all models atmos...

Page 2: ...HC AII 11 60 Oxygen Sensor AII 11 60 HC Oxygen Sensor BATT 1008 Battery 2x 1 5V AA BATT 1008 Battery 2x 1 5V AA P 1087 Instructions for Use P 1087 Instructions for Use A 1162 PCB Assy Main A 1162 PCB Assembly Main CABL 1006 Coil Cable TUBE 1019 7 32 OD Tubing 3 CC 1072 Carrying Case O2 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Page 3: ...e mail sales medical aii1 com web www aii1 com This manual may not be reproduced in whole or in part without the prior written consent of Analytical Industries Inc ISO 9001 2008 Certificate 485A AII 3000 AHC AII 3000 MHC AII 3000 M AII 3000 A 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Page 4: ...tanks intended for scuba diving for personnel safety area monitoring O2 deficiency in confined spaces checking breathing air tanks and confirming the O2 levels prior to welding Users must read the following statements as they are essential to re ducing the risk of use error due to ergonomic features of the device or the environment in which the device is intended to be used The devices have been d...

Page 5: ...enting the partial pressure of oxygen in the gas stream being analyzed The resulting oxygen reading is displayed by a large easy to read backlit liquid crystal display LCD that has a resolution of 0 1 oxygen The microprocessor is controlled from a keypad and provides features like system diagnostics warning indicators controls and an alarm capability for continu ous monitoring that enhance both sa...

Page 6: ...cluding the device is faulty If in doubt contact the manufacturer for assistance 4 NEVER operate the device in any manner described below doing so may compromise the clinical condition or the safety of patients users or other persons If the reading is unstable or a malfunction is suspected After the battery symbol appears in the LCD display Near equipment capable of emitting high levels of electro...

Page 7: ...ot to damage the case when removing the battery compartment cover 4 2 1 Procedure 1 Remove the device and the 2 AA 1 5V Alkaline batteries from the foam shipping container 2 Turn the device over so the shortest raised line on the battery compartment cover is pointing away from you 3 Lift the tripod wire stand up and away from the case 4 Grasp the case with both hands use your thumbs press down fir...

Page 8: ...0 AHC MHC When the HC hose connection version is ordered the device is shipped with the sensor installed 8 4 4 Controls 4 4 1 AII 3000 A AHC Oxygen Analyzers These analyzers employ a micro processor that is controlled by five 5 pushbuttons located on the keypad attached to front cover 1 ON OFF provides power to the electronics 2 ESCAPE aborts a previous selected option 3 ENTER selects a menu optio...

Page 9: ... A AHC Oxygen Analyzers The LCD alternately displays the following until the problem is corrected Corrective action 1 Expose the sensor to air or a gas containing approximately 20 9 oxygen 2 Connect or replace the cable connecting the sensor to the analyzer 3 Connect or replace the oxygen sensor 4 5 2 2 AII 3000 M MHC Oxygen Analyzers Performs the same routine and requires the same corrective acti...

Page 10: ...he LO alarm and proceed to the HI alarm or press the UP DOWN arrow keys to change the alarm set point 6 Press ENTER to save LO alarm set point and move to select the HI alarm 7 Press ENTER to skip the HI alarm and return to SAMPLING mode or press the UP DOWN arrow keys to change the alarm set point 8 Press ENTER to save HI alarm set point and return to SAMPLING mode 9 If no key is pressed within 5...

Page 11: ...ixty 60 seconds to allow the sensor to stabilize before the microprocessor takes the final reading 14 7 If the calibration is successful the display below left appears for three 3 seconds before defaulting to the display below right 8 The display above right requires a decision by the user refer to warnings at the beginning of section 4 7 to press ENTER and skip the 100 O2 calibration and return t...

Page 12: ...he sensor itself requires no maintenance and is simply replaced at the end of its useful life like a battery Inasmuch as the sensor is a transducer in its own right its ex pected life is not affected by whether the analyzer is ON or OFF The relationship between the sensor s signal and changes with the oxygen concentration is both proportional and linear thus allowing single point calibra tion Othe...

Page 13: ...he temperature and pressure altitude at which the ana lyzer will be operated 18 Effect of Humidity The analyzer is not affected by non condensing relative humidity RH Howev er the use of a humidifier to introduce water vapor and increase the moisture level of the gas mixture does affect the oxygen concentration and the resultant reading displayed by the analyzer The addition of water vapor increas...

Page 14: ...g displayed by the LCD If placing the entire sensor inside the controlled atmosphere review Section 5 2 Application Consideration Effect of Temperature 5 4 3 AII 3000 AHC and MHC Integral Oxygen Sensor AII 3000 AHC and MHC with their integral oxygen sensor requires connecting the tubing supplied section 4 2 1 above with the device to a hose barb attached to a pressure regulator controlling a sourc...

Page 15: ... simply find and register the male plug at the end of the coiled cable and insert it into the mating female jack at the rear of the sensor until it mates or snaps into place 3 Calibrate the device after replacing the oxygen sensor 6 3 2 Procedure AII 3000 AHC and AII 3000 MHC Integral Sensor 1 Tools required small bladed screwdriver 2 Place the device face down on a flat surface 3 Remove the two 2...

Page 16: ...ction Reading displayed by LCD drifts during calibration Wait 5 minutes and repeat calibration with sensor placed on flat surface not in your hand Check integrity of gas delivery system Check sensor s front o ring seal Verify calibration gas in not humidified Remove moisture covering sensor Replace sensor repeat calibration Analyzer reading climbs after calibration in 100 dry oxy gen when exposed ...

Page 17: ...ration Air or 100 oxygen after 8 hours of continuous use Compensation Temperature compensated Connections A M models 1x16mm thread HC models Tubing 1 4 Controls Soft touch keypad for ON OFF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 5 x 1 5 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing ...

Page 18: ... or in use The purpose of the Recommended Storage Life period is to ensure the user derives the Expected Life of 60 months 915 420 oxygen hours and does not lose the benefit of the warranty Warranty The 12 month 183 084 oxygen hours warranty period begins with ship ment from the factory and is limited to the first claim submitted is based on 8 1 Spare Parts Accessories AII 3000 A AII 3000 M AII 30...

Page 19: ...cidental or consequential losses or damages damages resulting from alterations misuse abuse lack of proper maintenance unau thorized repair or modification of the analyzer Service Contact us between 8 00am and 5 00pm PST Monday thru Thursday or before 12 00pm on Friday Trained technicians will assist you in diagnosing the prob lem and determining the appropriate course of action 31 10 Material Saf...

Reviews: