AII AII-3000 A Manual Download Page 15

22 

23 

Initially this procedure can be somewhat difficult. Care should be taken 
not to damage the case when removing the battery compartment cover. 

 

6.2.1  Procedure: 

1.  Turn  the  device  over  so  the 

shortest raised line on the battery 

compartment  cover  is  pointing 

away from you. 

2.  Lift  the  tripod  wire  stand  up  and 

away from the case. 

3.  Grasp  the  case  with  both  hands 

and  using  your  thumbs  press 

down firmly on the raised lines and push the battery compartment cover 
away from you. 

4.  Locate the positive (+) and negative (-) terminals on the battery. 
5.  Assure the battery contacts are clean. 
6. 

Align one battery’s positive (+) terminal with the corresponding (+) bat-

tery symbol molded into the case. 

7.  Insert the battery into the compartment. 
8.  Repeat with the remaining battery. 

9. 

Replace the battery compartment cover, make sure it snaps into position 

and is secured flush against the case. Replace the wire stand as required.

 

10. 

Calibrate the device after replacing the batteries.

 

 

6.3 Oxygen Sensor Replacement 

The design of the electronics is intended for only the Analytical Industries Inc. 
AII-11-60  or  AII-11-60-HC  Oxygen  Sensors.  Use  of  a  different  oxygen  sensor 

may result in an erroneous oxygen reading. 

 

NEVER -

 Open the oxygen sensor or probe the sensing surface, refer 

to Section

 

10 in the event the sensor should leak and someone comes 

in contact with the electrolyte from inside the sensor. 

 

6.3.1  Procedure AII-3000 A and AII-3000 M - External Sensor 

1.  Disconnect  the  cable  from  the  old  sensor  just  as  you  disconnect  a  tele-

phone jack from a wall plug. 

2.  To connect the new sensor simply find and register the male plug at the 

end of the coiled cable and insert it into the mating female jack at the rear 

of the sensor until it mates or snaps into place. 

3.  Calibrate the device after replacing the oxygen sensor. 

6.3.2  Procedure AII-3000 AHC and AII-3000 MHC - Integral Sensor 

1.  Tools required: small bladed screwdriver. 
2.  Place the device face down on a flat surface. 
3.  Remove  the  two  (2)  screws  from  the  upper  corners  of  the  rear  of  the 

device. 

4.  Move the tripod up, remove the battery compartment cover (see Battery 

Replacement) and remove the two (2) screws located on either side. 

5. 

Pull the rear section up ¼”

-

½”, turn it over and lay it next to the other 

section. 

6.  Locate the white connector at the end of the four (4) wires running from 

the sensor (the cylinder with the white label) to the top of the PCB. 

7.  With your left for finger and thumb, grasp the sides of the back end of the 

white connector where it is soldered to the PCB. 

8.  With your right fore finger and thumb, grasp the sides of the section of 

the white connector where the four (4) wires from the sensor terminate. 

9.  Separate the connector - hold the white connector section your left hand 

while  gently  pulling  and  wiggling  the  white  connector  section  with  your 

right hand until it unlocks. 

 
 
 
 
 
 
 
 
 
 
 
10.  The oxygen sensor inserts  into an adaptor (identified by a round  recess 

with  a  cylindrical  hose  adapter  in  the  center)  that  slides  into  grooves 

molded into the side of the case. 

11.  Hold  the  rear  section  of  the  case  down,  grasp  the  square  edges  of  the 

adaptor, lift up (lift straight up so as not to strip the grooves molded into 

the adaptor and case)  and remove the adaptor and oxygen sensor as a 
single component. 

 

姚经理19961810948(微信同号)

QQ:2316517588

邮箱:[email protected]

Summary of Contents for AII-3000 A

Page 1: ...FF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 75 x 1 375 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing 0 95 RH LED Indicators A models None M models upon activation of alarms Linearity 1 under constant conditions Pressure Inlet A M models ambient HC models regulate Vent all models atmos...

Page 2: ...HC AII 11 60 Oxygen Sensor AII 11 60 HC Oxygen Sensor BATT 1008 Battery 2x 1 5V AA BATT 1008 Battery 2x 1 5V AA P 1087 Instructions for Use P 1087 Instructions for Use A 1162 PCB Assy Main A 1162 PCB Assembly Main CABL 1006 Coil Cable TUBE 1019 7 32 OD Tubing 3 CC 1072 Carrying Case O2 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Page 3: ...e mail sales medical aii1 com web www aii1 com This manual may not be reproduced in whole or in part without the prior written consent of Analytical Industries Inc ISO 9001 2008 Certificate 485A AII 3000 AHC AII 3000 MHC AII 3000 M AII 3000 A 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Page 4: ...tanks intended for scuba diving for personnel safety area monitoring O2 deficiency in confined spaces checking breathing air tanks and confirming the O2 levels prior to welding Users must read the following statements as they are essential to re ducing the risk of use error due to ergonomic features of the device or the environment in which the device is intended to be used The devices have been d...

Page 5: ...enting the partial pressure of oxygen in the gas stream being analyzed The resulting oxygen reading is displayed by a large easy to read backlit liquid crystal display LCD that has a resolution of 0 1 oxygen The microprocessor is controlled from a keypad and provides features like system diagnostics warning indicators controls and an alarm capability for continu ous monitoring that enhance both sa...

Page 6: ...cluding the device is faulty If in doubt contact the manufacturer for assistance 4 NEVER operate the device in any manner described below doing so may compromise the clinical condition or the safety of patients users or other persons If the reading is unstable or a malfunction is suspected After the battery symbol appears in the LCD display Near equipment capable of emitting high levels of electro...

Page 7: ...ot to damage the case when removing the battery compartment cover 4 2 1 Procedure 1 Remove the device and the 2 AA 1 5V Alkaline batteries from the foam shipping container 2 Turn the device over so the shortest raised line on the battery compartment cover is pointing away from you 3 Lift the tripod wire stand up and away from the case 4 Grasp the case with both hands use your thumbs press down fir...

Page 8: ...0 AHC MHC When the HC hose connection version is ordered the device is shipped with the sensor installed 8 4 4 Controls 4 4 1 AII 3000 A AHC Oxygen Analyzers These analyzers employ a micro processor that is controlled by five 5 pushbuttons located on the keypad attached to front cover 1 ON OFF provides power to the electronics 2 ESCAPE aborts a previous selected option 3 ENTER selects a menu optio...

Page 9: ... A AHC Oxygen Analyzers The LCD alternately displays the following until the problem is corrected Corrective action 1 Expose the sensor to air or a gas containing approximately 20 9 oxygen 2 Connect or replace the cable connecting the sensor to the analyzer 3 Connect or replace the oxygen sensor 4 5 2 2 AII 3000 M MHC Oxygen Analyzers Performs the same routine and requires the same corrective acti...

Page 10: ...he LO alarm and proceed to the HI alarm or press the UP DOWN arrow keys to change the alarm set point 6 Press ENTER to save LO alarm set point and move to select the HI alarm 7 Press ENTER to skip the HI alarm and return to SAMPLING mode or press the UP DOWN arrow keys to change the alarm set point 8 Press ENTER to save HI alarm set point and return to SAMPLING mode 9 If no key is pressed within 5...

Page 11: ...ixty 60 seconds to allow the sensor to stabilize before the microprocessor takes the final reading 14 7 If the calibration is successful the display below left appears for three 3 seconds before defaulting to the display below right 8 The display above right requires a decision by the user refer to warnings at the beginning of section 4 7 to press ENTER and skip the 100 O2 calibration and return t...

Page 12: ...he sensor itself requires no maintenance and is simply replaced at the end of its useful life like a battery Inasmuch as the sensor is a transducer in its own right its ex pected life is not affected by whether the analyzer is ON or OFF The relationship between the sensor s signal and changes with the oxygen concentration is both proportional and linear thus allowing single point calibra tion Othe...

Page 13: ...he temperature and pressure altitude at which the ana lyzer will be operated 18 Effect of Humidity The analyzer is not affected by non condensing relative humidity RH Howev er the use of a humidifier to introduce water vapor and increase the moisture level of the gas mixture does affect the oxygen concentration and the resultant reading displayed by the analyzer The addition of water vapor increas...

Page 14: ...g displayed by the LCD If placing the entire sensor inside the controlled atmosphere review Section 5 2 Application Consideration Effect of Temperature 5 4 3 AII 3000 AHC and MHC Integral Oxygen Sensor AII 3000 AHC and MHC with their integral oxygen sensor requires connecting the tubing supplied section 4 2 1 above with the device to a hose barb attached to a pressure regulator controlling a sourc...

Page 15: ... simply find and register the male plug at the end of the coiled cable and insert it into the mating female jack at the rear of the sensor until it mates or snaps into place 3 Calibrate the device after replacing the oxygen sensor 6 3 2 Procedure AII 3000 AHC and AII 3000 MHC Integral Sensor 1 Tools required small bladed screwdriver 2 Place the device face down on a flat surface 3 Remove the two 2...

Page 16: ...ction Reading displayed by LCD drifts during calibration Wait 5 minutes and repeat calibration with sensor placed on flat surface not in your hand Check integrity of gas delivery system Check sensor s front o ring seal Verify calibration gas in not humidified Remove moisture covering sensor Replace sensor repeat calibration Analyzer reading climbs after calibration in 100 dry oxy gen when exposed ...

Page 17: ...ration Air or 100 oxygen after 8 hours of continuous use Compensation Temperature compensated Connections A M models 1x16mm thread HC models Tubing 1 4 Controls Soft touch keypad for ON OFF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 5 x 1 5 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing ...

Page 18: ... or in use The purpose of the Recommended Storage Life period is to ensure the user derives the Expected Life of 60 months 915 420 oxygen hours and does not lose the benefit of the warranty Warranty The 12 month 183 084 oxygen hours warranty period begins with ship ment from the factory and is limited to the first claim submitted is based on 8 1 Spare Parts Accessories AII 3000 A AII 3000 M AII 30...

Page 19: ...cidental or consequential losses or damages damages resulting from alterations misuse abuse lack of proper maintenance unau thorized repair or modification of the analyzer Service Contact us between 8 00am and 5 00pm PST Monday thru Thursday or before 12 00pm on Friday Trained technicians will assist you in diagnosing the prob lem and determining the appropriate course of action 31 10 Material Saf...

Reviews: