Measuring rate function
: for input (see Sec.5.5)
Measuring rate : 10
Continuous measuring point scan (basic setting) :
Cont :
"
Continuous storage :
Save : -
Activate continuous storage :
<ON>
"
Continuous output :
Output : -
Activate continuous output :
<ON>
"
At the conversion rate of 50 measuring operations per second, the shorter
evaluation times lead to the following restrictions :
1. The increased conversion rate only takes effect after a measuring op-
eration has been initiated; before this, the device remains set at 10 meas-
uring operations per second.
2. During measuring operations at a higher measuring rate, the ALMEMO
connectors are no longer monitored; i.e. the connector configuration can
only be modified when the measuring device has stopped.
3.Alarm value printout is not provided.
4.The analog output is not used.
5.At rates above 10 measuring operations per second, mains hum sup-
pression is not provided; as a result, accuracy may be adversely affected
by interference over the connection lines; (wherever possible use twisted
wires !)
7.5 Time and Date of Start, Time and Date of End
A sequence of measurements can, at certain points in time, be automatically
started and stopped. For this purpose the start time and date and the end time
and date can be programmed. If a date has not been specified the
measurement will be performed every day in the defined period. The current
time must be programmed beforehand.
Function ´Start Time´:
StartTime: 07:00:00
Function ´End Time´:
EndTime: --------
Input (see 5.5) in format:
hh:mm:ss
Function ´Start Date´:
StartDate:01.05.00
Function ´End Date´:
EndDate: --------
Input (see 5.5) in format:
dd:mm:yy
To clear values, select function and then use:
<
CLR
>
If the start time of a measurement has been
programmed, the status line will include the symbol:
´
´
!
If the end time of a measurement has been
programmed, the status line will include the symbol:
´
´
!
ALMEMO
®
2590-9
43
Times and Cycles