XGS-600 Gauge Controller
72
DRAFT 4/
15/14
Setpoint Commands
Read Setpoint States
#aa03
>00hh
where hh is a hex value
where each bit
represents
setpoints 1 - 8 and
value 0=Off, 1=On
Setpoint numbers
correspond to the
numbers on the
Setpoint screen.
Read assigned setpoints #aa04cn
>00hh
where hh is a hex value
where each bit
represents
setpoints 1 - 8 and
value 0=Not assigned,
1=Assigned
Set Setpoint
OFF/ON/Auto
#aa5Ehx
where h is setpoint
number 1-8 and x is
state:
0 = OFF (manual
override)
1 = ON (manual
override)
3 = Auto
(based on pressure)
Example: #005E83 sets
setpoint #8 to Auto.
>
Read Setpoint
OFF/ON/Auto
#aa5Fh
where h is setpoint
number 1-8
>x
where x is state:
0 = OFF (manual
override)
1 = ON (manual
override)
3 = Auto (based on
pressure)
Set Setpoint On
pressure level
#aa6hcnx.xxxE-xx
where h is setpoint
number 1-8
>
Table B-1 Serial Command Set (Continued)
Description
Command
Response from
XGS-600
Notes