XGS-600 Gauge Controller
70
DRAFT 4/
15/14
Ion Gauge Commands (HFIG and IMG gauges)
Assign Tube Type to
Sensor
(only HFIG and IMG
boards)
#aa16cnxx
where xx =
11 for IMG100 (default)
13 for IMG300
51 for MBA100/MBA2
52 for MBA200
63 for 563
64 for 564
71 for 571
72 for 572
80 for UHV24 (default)
81 for UHV24p
>
Read Sensor Tube Type
(only HFIG and IMG
boards)
#aa17cn
>xx where xx =
11 for IMG100 (default)
13 for IMG300
51 for MBA100/MBA2
52 for MBA200
63 for 563
64 for 564
71 for 571
72 for 572
80 for UHV24 (default)
81 for UHV24p
Set Emission OFF
#aa30cn
>
Set Emission ON (Fil 1) #aa31cn
>
Also used for
single-filament gauges.
Read Emission status
#aa32cn
>xx
where:
xx = 00 is OFF
xx = 01 is ON
Set Emission ON (Fil2)
#aa33cn
>
Read Filament Lit
#aa34cn
>xx
where:
xx = 01 is Filament 1
xx = 02 is Filament 2
Set Auto Fil Advance
Off
#aa35
>
Set Auto Fil Advance
On
#aa36 >
Table B-1 Serial Command Set (Continued)
Description
Command
Response from
XGS-600
Notes