XGS-600 Gauge Controller
56
DRAFT 4/
15/14
Gauge Connector
❑
Collector: SMB coax
❑
Fil and Grid:AMP 5 pos Univ Mate-N-Lok
Remote I/O
❑
Connector: 6 pin IEEE 1394 (Firewire type)
❑
Inputs for Emission on/off and DEGAS on/off are user select-
able to be either Active LO or Active HI. Driving the input to
the active state turns on that function.
❑
HI = +5 V max, 2.4 V min
❑
LO = 0.8 V max, 0 V min
❑
Filament Select input: Pin 1 open or driven HI selects Fil1.
Pin 1 grounded selects Fil2.
❑
Status Output goes HI during error condition
❑
Analog Output
Allowable Cable Length
❑
up to 50’ using Agilent standard cable
❑
up to 300’ using under 10’ Agilent standard cable + Agilent
extension cable
Table A-2 HFIG Gauge Specifications
Item
Description
Summary of Contents for XGS-600
Page 2: ...Copyright 2014 Agilent Technologies XGS 600 Gauge Controller ...