This menu contains parameters that determine the operating mode of
the inputs and outputs of the controller.
START MODE
: it distinguish three ways to start/stop the pump
according to following description:
-
START INTERLOCK, the controller checks for interlock mode;
see interlock type description for further details.
-
START/STOP LEVEL; if selected, close the START/STOP contact
on I/O connector (1-9 and 2-15) to start the pump and close the
INTERLOCK signal (3-9 and 4-15) to stop the pump.
INTERLOCK TYPE:
selects the interlock mode on I/O connector:
-
CONTINOUS: if enabled, the unit wants the contact on I/O
connector to be closed; should be the contact open, error
condition WAITING FOR INTERLOCK is displayed;
-
IMPULSIVE: if enabled, the unit wants the interlock to be closed
only at start command; should be removed the contact after start
command or during normal operation, no error is displayed.
ANALOG OUTPUT TYPE
: referred to the ANALOG OUTPUT SIGNAL
on I/O connector; it can be related to one of following quantity:
-
FREQUENCY, default value
-
POWER
-
TEMPERATURE
SPEED DISPLAY
: the rotational speed can be displayed in
-
KRPM
-
Hz
LCD CONTRAST
, increases/decreases the contrast on LCD display.