Technical Information
Window Meanings
17
Turbo-V 1001 Rack Controller User Manual / 87-901-033-01
241/288
N.
Read/
Write
Data
Type
Description
Admitted Values
167 R/W
L
Stop Speed Reading (when
active, the pump rotation
speed is measured also in
Braking status)
0 = No
1 = Yes
(default = 1)
200 R
N
Pump
current expressed in mA dc
201 R
N
3 phase voltage expressed in
V
rms
202 R
N
Output power expressed in W
203 R
N
Driving frequency expressed
in Hz
204 R
N
Pump bearing temperature
expressed
in °C
205 R
N
System status
Stop = 0
Waiting intlk = 1
Ramp-Up = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
206 R
N
Error code
Bit description: see the
following figure
211 R
N
Controller Internal temperature
in °C
222 R
N
Body pump temperature in °C
232 R
N
Rotational frequency in Hz
300 R
N
Cycle time in minutes (zeroed
by the reset command)
0 to 999999
301 R
N
Cycle number (zeroed by the
reset command)
0 to 999999
302 R
N
Pump life in hours (zeroed by
the reset command)
0 to 999999