Agilent U1251A/U1252A User’s and Service Guide
7
Performance Tests and Calibration
Step
Test Function
Range
5520A Output
Error from nominal 1 year
U1251A
U1252A
4
Turn the rotary switch to
V position (for model U1252A), to
V position (for model
U1251A)
5 V
5 V
± 2 mV
±1.75 mV
50 V
50 V
± 20 mV
± 17.5 mV
500 V
500 V
± 200 mV
± 200 mV
1000 V
1000 V
± 800 mV
± 800 mV
5
Press
button to go to the
V mode
[1]
5 V
5 V,1 kHz
N/A
± 22.5 mV
5 V, 10 kHz
N/A
± 22.5 mV
5 V, 20 kHz
N/A
± 41.5 mV
5 V, 100 kHz
N/A
± 187 mV
50 V
50 V, 1 kHz
N/A
± 225 mV
50 V, 10 kHz
N/A
± 225 mV
50 V, 20 kHz
N/A
± 415 mV
50 V, 100 kHz
N/A
± 1.87 V
500 V
500 V, 1 kHz
N/A
± 2.25 V
500 V, 10 kHz
N/A
± 2.25 V
1000 V
1000 V, 1 kHz
N/A
± 8.0 V