Performance Tests and Calibration
7
Agilent U1251A/U1252A User’s and Service Guide
133
Step
Test Function
Range
5520A Output
Error from nominal 1 year
U1251A
U1252A
6
Turn the rotary switch to the
mV position
50 mV
50 mV
± 75
μ
V
[2]
± 75
μ
V
[2]
500 mV
500 mV
± 0.2 mV
± 0.175 mV
–500 mV
± 0.2 mV
± 0.175 mV
1000 mV
1000 mV
± 0.8 mV
± 0.75 mV
–1000 mV
± 0.8 mV
± 0.75 mV
7
Press
button to go to
the
mV
mode
[1]
50 mV
50 mV, 1 kHz
± 0.34 mV
± 0.24 mV
50 mV, 10 kHz
± 0.54 mV
± 0.39 mV
50 mV, 20 kHz
N/A
± 0.415 mV
50 mV, 30 kHz
± 0.86 mV
N/A
50 mV, 100 kHz
N/A
± 1.87 mV
500 mV
500 mV, 45 Hz
± 5.6 mV
± 8.1 mV
500 mV, 1 kHz
± 3.25 mV
± 2.25 mV
500 mV, 10 kHz
± 5.4 mV
± 2.25 mV
500 mV, 20 kHz
N/A
± 4.15 mV
500 mV, 30 kHz
± 8.6 mV
N/A
500 mV, 100 kHz
N/A
± 18.7 mV
1000 mV
1000 mV, 1 kHz
± 8.5 mV
± 6.5 mV
1000 mV, 10 kHz
± 12.5 mV
± 6.5 mV
1000 mV, 20 kHz
N/A
± 11.5 mV
1000 mV, 30 kHz
± 20.0 mV
N/A
1000 mV, 100 kHz
N/A
± 47.0 mV