2-4
TS-8900 Functional Test System Site Preparation and Installation Guide
2
Safety and Regulatory Information
Before Applying Power
Verify that the product is set to match the available line voltage
and all safety precautions are taken. Note the external markings
of the instruments described in
”.
Operator Safety Information
Cables and connectors are considered inaccessible if a tool (e.g.
screwdriver, wrench, socket, etc.) or a key (equipment in a
locked cabinet) is required to gain access to a conductive
surface connected to any cable conductor (High, Low or Guard).
Verify there are multiple and sufficient protective means (rated
for the voltages you are applying) to assure the operator will
NOT come into contact with any energized conductor even if
one of the protective means fails to work as intended.
For example, the inner side of a case, cabinet, door cover or
panel can be covered with an insulating material as well as
routing the test cables to the front panel connectors of the
module through non-conductive, flexible conduit such as that
used in electrical power distribution.
WA R N I N G
M
Module connectors and Test Signal cables connected to them
cannot be operator accessible.
WA R N I N G
Assure the equipment under test has adequate insulation
between the cable connections and any operator-accessible parts
(doors, covers, panels shields, cases, cabinets, etc.)