3
Basic Operation
Set alarm (temperature tolerance range)
30
Seahorse XFe Analyzer Operating Manual
2
Ensure your ambient conditions support the desired target temperature
(12-20 °C above ambient). (See the temperature chart shown in
page 28.)
3
Other Temperature Widget functions are:
•
Turn the heater ON/OFF.
•
Set the tolerance range for temperature fluctuation. If the temperature is
above or below the acceptable tolerance range from the temperature set
point, the Temperature Widget will change color, and the Status Indicator
light (top of the XFe Analyzer) will change from blue to amber. For
networked XFe Controllers, Wave Controller software automatically sends
an email notification to specified recipients.
4
To save any changes on the Tray Temperature window, click
Save
.
Set alarm (temperature tolerance range)
To set the alarm:
1
Check the
Alarm On
check box in the Tray Temperature window.
(See
2
Click
Save
.
To disable the alarm, uncheck the
Alarm On
box, then click
Save
.
If the Tray Temperature exceeds the Tolerance Range and the alarm is activated,
click
Reset Alarm
to acknowledge and reset the Tray Temperature alarm.
To ensure the Tray Temperature starts within the Tolerance Range, check the
current temperature of the XFe Analyzer before beginning an assay. For any
suspected temperature issues or unexpected temperature fluctuations, contact
Technical Support. (See
NOTE
Changing the Target Temperature requires OVERNIGHT equilibration to the
new set point.
Summary of Contents for Seahorse XFe
Page 1: ...Seahorse XFe Analyzer Operating Manual...
Page 4: ...4 Seahorse XFe Analyzer Operating Manual...
Page 12: ...1 Introduction Technical Specifications 12 Seahorse XFe Analyzer Operating Manual...
Page 38: ...4 Maintenance Additional Resources 38 Seahorse XFe Analyzer Operating Manual...
Page 39: ......