RS-232 and RS-485 Options for the senTorr Gauge Controller
1-10
DRA
FT
Read degas status
#aa42l1tt
>hht
where:
❑
hh = 00 is OFF
❑
hh = 01 is ON
Read gas correction
#aa50l1t
>x.xxxt
Set gas correction
#aa51l1x.xxxt
>t
Read Emission current
#aa52l1t
>x.xxxt
Set Emission current
#aa53l1x.xxxt
>t
Read Sensitivity
#aa54l1t
>xx.xxt
Set Sensitivity
#aa55l1xx.xxt
>t
Set Setpoint pressure level
#aa6hcnx.xxxE-xxt
where h is the setpoint relay
number, 1 – 4
>t
Set Setpoint hysteresis level
#aa7hcnx.xxxE-xxt
where h is the setpoint relay
number, 1 – 4
>t
Read Setpoint pressure level
#aa8ht
where h is the setpoint number,
1 – 4
>x.xxxE-xxt
Read Setpoint hysteresis level
#aa9ht
where h is the set point number,
1 – 4
>x.xxxE-xxt
Set Thermocouple Cal
#aaA1Tnt
>t
Set Thermocouple Update Rate
to Slow (standard)
#aaA7t
>t
Set Thermocouple Update Rate
to Fast
#aaA8t
>t
Read Thermocouple Update Rate #aaA9t
>hht
hh = 00 is Slow
hh = 01 is Fast
Set Auto-On
#aaB0l1T1x.xE-xxt
>t
Read Auto-On
#aaB1l1t
>T1x.xExxt
Table 1-8
Serial Command Set (Continued)
Function
Command
Response