RS-232 and RS-485 Options for the senTorr Gauge Controller
1-8
DRA
FT
Table 1-8 lists the senTorr serial commands. All lower case characters must be replaced as
follows:
aa
unique 2 digit hexadecimal bus address (""00" ...:99"), as set on the
unit
hh
2 character hex data value (00 ...FF")
c
1 character channel type ("1", 'or") where I = BNCC
T=TC
n
1 character channel number where IG = 1, TC1 = 1, TC2 = 2 x = 1
character data value ("1" '19")
t
message terminator character (#13, <carriage return>
Table 1-8
Serial Command Set
Function
Command
Response
Read senTorr configuration
#aa01t
>hhhhhhhhhht
where the card id codes are:
❑
20 = BA configured for
Broad-range gauges
❑
30 = BA configured for
standard Bayard-Alpert
gauges
❑
38 = Cold cathode
❑
40 = Thermocouple board
❑
50 = Setpoint
❑
FE = Empty slot
Read gauge pressure
#aa02cnt
>x.xxxE-xxt
Read Setpoint State
#aa03t
>00hht
where hh bits 0 – 3represent
setpoints 1 – 4 and value is state
(0 = off, 1 = on)
Read programmed setpoints
#aa04cnt
>00hht
Where bits 0 – 3 represent
setpoints 1 – 4 and value is
assignment (0 = not assigned, 1
= assigned)
Read software revision
#aa05t
>hhhht
where hh = 0 – 9 and the
revision is hh.hh r