4-24
Service Guide N5242-90001
Troubleshooting
PNA Series Microwave Network Analyzers
Measurement System Troubleshooting
N5242A
EEPROM Headers
The network analyzer application uses the firmware revision information stored in the pc
board header EEPROM. If the information stored in any EEPROM is incorrect, the
network analyzer may not operate properly.
The following table lists the pc boards in your network analyzer that contain EEPROM
headers. The pc boards are listed by name and part number and the correct firmware
revision code is given for each.
To view this EEPROM header information on the network analyzer display: press
UTILITY
, then
, then
, then
.
If the information is incorrect for any of the pc boards, contact Agilent Technologies. Refer
to
“Contacting Agilent” on page 2-13
Error Messages
SOURCE UNLEVELED:
The source ALC circuit on the A19 test set motherboard is running
open-loop. Check the cable connections for the A21 HMA26.5 and the A19 test set
motherboard.
PC Board Name
PC Board Part
Number
Memory
Type ID
Hardware
ID
Firmware
Revision
System
Service
Utilities
View EEPROM Headers
Summary of Contents for N5242A
Page 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Page 33: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Page 75: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Page 119: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Page 153: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Page 237: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Page 303: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...