7-58
Service Guide N5242-90001
Repair and Replacement Procedures
PNA Series Microwave Network Analyzers
Removing and Replacing the Front Panel LED Boards
N5242A
Removing and Replacing the Front Panel LED Boards
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16 inch open-end torque wrench (set to 10 in-lb)
• 1 inch open-end torque wrench (set to 72 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
for this procedure.
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing
of the cables is necessary to remove the assembly. However, do not over-bend
the semirigid cables.
1. Disconnect the power cord.
2. Remove the test port couplers. Refer to
“Removing and Replacing the A29–A32 Test
3. Remove two screws (item
①
) from the LED board to be removed and remove the LED
board from the analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
Summary of Contents for N5242A
Page 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Page 33: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Page 75: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Page 119: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Page 153: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Page 237: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Page 303: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...