7-6
Service Guide N5242-90001
Repair and Replacement Procedures
PNA Series Microwave Network Analyzers
Removing the Covers
N5242A
Removing the Covers
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
Removing the Outer Cover
CAUTION
This procedure is best performed with the analyzer resting on its front
handles in the vertical position.
Do not place the analyzer on its front panel
without the handles.
This will damage the front panel assemblies.
Refer to
for this procedure.
1. Disconnect the power cord.
2. Remove the strap handles (item
①
) by loosening the screws (item
②
), with a T-20 TORX
driver, on both ends until the handle is free of the analyzer.
3. Remove the foot locks (item
③
) from the four bottom feet (item
④
) and then remove the
four bottom feet from the outer cover.
4. Remove the four rear panel feet (item
⑤
) by removing the center screws (item
⑥
with a
T-20 TORX driver,.
5. Slide the outer cover toward the rear of the analyzer and remove it.
Removing the Inner Cover
Refer to
for this procedure.
1. With a T-10 TORX driver, remove the 12 pan head screws (item
⑦
).
2. With a T-10 TORX driver, remove the 9 flat head screws (item
⑧
).
3. Lift off the cover.
Replacement Procedure
Reverse the order of the removal procedures.
Summary of Contents for N5242A
Page 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Page 33: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Page 75: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Page 119: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Page 153: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Page 237: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Page 303: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...