3-2
Adjustments
Introduction
Introduction
This chapter provides adjustment procedures for the Agilent 86100A/B/C Infiniium DCA Main-
frame. Equipment required for individual adjustments is listed in the adjustment descriptions in
this chapter. Equipment satisfying the critical specifications listed may be substituted for the rec-
ommended model.
This chapter documents both manual and automated testing. The adjustments can be performed
either manually or with the VISION test system and the Agilent DSA Test Executive.
N O T E
Do a Performance Verification before attempting any adjustments for customers requiring “As
Received” or “Pre Cal” data.
C A U T I O N
Avoid damage to plug-in front panel connectors. Use 2.4 mm and 3.5 mm connector savers. These
connector savers are a supplied accessory. Minimize connector swapping during the procedures
to avoid connector wear. All connectors on test tools and adapters should be inspected both
visually and mechanically every few calibrations.
C A U T I O N
All connectors should be clean and undamaged to ensure accurate measurements. All 2.4 mm and
3.5 mm connectors should be mechanically and visually checked before inserting any calibration
test tool into them. Damaged connectors or loose connectors may cause the performance
verification tests to fail.
C A U T I O N
Avoid sharp bends in 2.4 mm, 3.5 mm, SMA, and optical cables. When mating 2.4 mm to 2.4 mm
or 3.5 mm to 3.5 mm, torque all connections to 8 in/lbs. When mating 3.5 mm to SMA or SMA
to SMA, torque all connections to 5 in/lbs.
C A U T I O N
The module inputs are very sensitive to static discharge. Failure to observe proper antistatic
procedures may damage the gallium arsenide samplers. ESD damage is not covered under the
warranty. All maintenance or operation should be performed with an antistatic mat and wrist strap.
Refer to “Electrostatic Discharge Information” on page 1-8
for further information.
C A U T I O N
Electrostatic discharge can seriously damage the module’s electrical inputs. To eliminate any
electrostatic build up from a cable you’re connecting to the module, connect a female short to
either end of the cable. Touch the short to an input connector hex nut on the module to discharge
any static build up to ground. Remove the short. Use this procedure for all cables before
connecting them to the module.
Summary of Contents for Infinium DCA 86100A
Page 1: ...Agilent 86100A B C Wide Bandwidth Oscilloscope Mainframe Service Guide...
Page 6: ...Contents 4...
Page 28: ...1 22 General Information DCA Options and Option Upgrades...
Page 31: ...2 3 Automated Testing and the VISION Test System Introduction Figure 2 1 Equipment Rack Layout...
Page 110: ...5 14 Instrument Refurbishment Refurbishing an 86100A B...
Page 132: ...6 22 Instrument Modifications A13 Acquisition Board PLD Header Modification 86100A Only...
Page 189: ...8 7 Replaceable Parts 86100A Front View Identification...
Page 191: ...8 9 Replaceable Parts 86100A Front Inside Panel Identification...
Page 193: ...8 11 Replaceable Parts 86100A Front View Front Panel Removed Identification...
Page 195: ...8 13 Replaceable Parts 86100A Rear View Identification...
Page 197: ...8 15 Replaceable Parts 86100A Left and Right Side Identification...
Page 199: ...8 17 Replaceable Parts 86100A Left Side Cover Removed Identification...
Page 201: ...8 19 Replaceable Parts 86100A Right Side Cover Removed Identification...
Page 203: ...8 21 Replaceable Parts 86100A Bottom View Identification 4 places 2 1 4 3...
Page 205: ...8 23 Replaceable Parts 86100A Bottom View Cover Removed Identification...
Page 207: ...8 25 Replaceable Parts 86100A Top View Cover Removed Identification...
Page 210: ...8 28 Replaceable Parts 86100A Block Diagrams 86100A Block Diagram 1 of 2...
Page 211: ...8 29 Replaceable Parts 86100A Block Diagrams 86100A Block Diagram 2 of 2...
Page 212: ...8 30 Replaceable Parts 86100A Block Diagrams 86100A Option 001 Block Diagram 1 of 2...
Page 213: ...8 31 Replaceable Parts 86100A Block Diagrams 86100A Option 001 Block Diagram 2 of 2...
Page 214: ...8 32 Replaceable Parts 86100A Block Diagrams...
Page 221: ...9 7 Replaceable Parts 86100B Front View Identification...
Page 223: ...9 9 Replaceable Parts 86100B Front Inside Panel Identification...
Page 225: ...9 11 Replaceable Parts 86100B Front View Front Panel Removed Identification...
Page 233: ...9 19 Replaceable Parts 86100B Top View Cover Removed Identification...
Page 234: ...9 20 Replaceable Parts 86100B Rear Panel Identification Rear Panel Identification...
Page 236: ...9 22 Replaceable Parts 86100B Block Diagrams 86100B Block Diagram 1 of 2...
Page 237: ...9 23 Replaceable Parts 86100B Block Diagrams 86100B Block Diagram 2 of 2...
Page 238: ...9 24 Replaceable Parts 86100B Block Diagrams 86100B Option 001 Block Diagram 1 of 2...
Page 239: ...9 25 Replaceable Parts 86100B Block Diagrams 86100B Option 001 Block Diagram 2 of 2...
Page 240: ...9 26 Replaceable Parts 86100B Block Diagrams...
Page 247: ...10 7 Replaceable Parts 86100C Major Assembly and Cable Location...
Page 249: ...10 9 Replaceable Parts 86100C Top View Cover Removed Identification...
Page 251: ...10 11 Replaceable Parts 86100C Front View Identification...
Page 255: ...10 15 Replaceable Parts 86100C Rear Panel Identification Rear Panel Identification...
Page 261: ...10 21 Replaceable Parts 86100C Block Diagrams...
Page 262: ...10 22 Replaceable Parts 86100C Block Diagrams 86100C Block Diagram 1 of 2...
Page 263: ...10 23 Replaceable Parts 86100C Block Diagrams 86100C Block Diagram 2 of 2...
Page 264: ...10 24 Replaceable Parts 86100C Block Diagrams 86100C Option 001 Block Diagram 1 of 2...
Page 265: ...10 25 Replaceable Parts 86100C Block Diagrams 86100C Option 001 Block Diagram 2 of 2...
Page 266: ...10 26 Replaceable Parts 86100C Block Diagrams...