G4208A User’s Guide
148
5
Maintenance and Repair
Firmware Updates
Upgrade from A.05.11 to A.05.1x
If the Instant Pilot needs to be updated with a new firmware revision , follow
these steps:
1
Disconnect all 1100/1200 series module from that module that hosts the
LAN/RS- 232 interface.
2
Upgrade this module to firmware A.06.xx first (otherwise the Instant Pilot is
not visible in the LAN/RS- 232 Update Tool 2.10).
3
After boot, enter the Maintenance screen and press the
Controller
button.
4
Press the
Transfer
button.
The Instant Pilot will switch into resident mode (black screen) and the Update
Tool lists now the Instant Pilot.
5
Select the Instant Pilot for update to the target firmware revision
6
Start the update.
7
After boot of the Instant Pilot,
•
the Instant Pilot shows “Scanning System ...”
•
the LAN/RS- 232 Update Tool 2.10 indicates “100% updated” but does not
display the result dialog.
8
Press
Cancel
(red cross button) on the LAN/RS- 232 Update Tool 2.10 to stop
the Update process.
The Instant Pilot is no longer listed in the tree view of the Update Tool.
9
Downgrade the 1100/1200 module to the target firmware (A.05.11/13 or
A.05.09/10).
10
When finished,
•
the LAN/RS- 232 Update Tool 2.10 shows all 1100/1200 modules without
the Instant Pilot.
•
the Instant Pilot shows all modules of the system in the Welcome screen.
11
Close the connection and the LAN/RS- 232 Update Tool 2.10.